Page 167 - Reciprocating Compressors Operation Maintenance
P. 167

Operation and Maintenance  of Reciprocating  Compressors  1  53


            Inertia  may  close  it  too  far,  even  back  to  the  seat  before  flow
          forces  cause  it to open  again. By the  end  of  the piston  stroke,  it may
          have oscillated  several times.
            Furthermore,  at  the  end  of  the  stroke,  inertia  may  carry  the  valve
          open  instead  of  back  toward  the  seat,  and  when  backwash finally
          closes the valve, it is with a slam.
            Examination  of  the  valve  and  seat  may  confirm fluttering. If  the
          stop  plate  shows  no markings, it  may indicate  that the  valve has  not
          fully  opened.  Normally the  back  of  the  valve plate  itself  will  have
          some  sort  of  pattern from  the  springs.  In  a circular or  plate  valve, if
          there is no definite pattern, and the plate appears  to spin, then flutter-
          ing is likely to cause failure.
            The  solution to  fluttering  is  to  lower  the  lift  and/or  to  use  lighter
          springs.
         • Resonance  or  pulsations can  upset  normal  valve action.  The  ampli-
          tude of pulsations is of considerably less consequence than the phase
          relation  of the pressure  wave and crank angle, A pulsation can  cause
          late closing, thus slamming, as previously mentioned.
            A  series  of  high  peak-to-peak  pressure  fluctuations  surround  the
          valve, usually as the result of an improperly  designed piping  system.
          As  the sealing  element  tries to follow this pressure  change, it flutters
          erratically.
            In  some cases, the sealing  elements  will pound against the seat and
          guard  many  times  during  a  single  piston  stroke,  causing  breakage
          from  impact fatigue.
            Pulsations  of  relatively  high  frequency,  for  example,  between  50
          and  100 Hz, are detrimental to compressor  valve life.
            Figure 3-19 depicts  on a time basis what happens.  The valve is not
          only  open  at dead  center, but the rate of pressure  change  is high. The
          valve is slammed shut by back flow  after dead  center.
            The  problem  can be  solved  by  eliminating pulsation  through pip-
          ing  changes.  This  is  a  case  where  the  problem  is  outside  the  valve
          designer's control and valve changes  will not correct  the problem.
         • Flow  pattern  failures are  infrequent, but  they  have happened.  There
          may  be a disturbance within the  cylinder passage  leading to a valve
          that causes  some type of erratic valve action.
            The  cure  is  to change  the  lift  and/or  springing (depending on  the
          type of failure), or even to change the valve type.
   162   163   164   165   166   167   168   169   170   171   172