Page 100 - Tandem Techniques
P. 100

Page 81

            Thus (a) can be controlled by varying (i). The rotational resolution (Da) for a controlled change in
            current through the coils of (Di) can be as little as 10  but due to heat losses in the coil, the maximum
                                                               -5
            value of (Da) is limited to about ± 2°. A diagram of the detector is shown in Figure 22.










































                                                         Figure 2.23
                                           A Chiral Detector Monitoring Optical Rotation
                                                   Courtesy of JM Science Inc.

            Light from a tungsten lamp passes through two condenser lenses to a polarizer. Plane polarized light
            from the polarizer passes through a temperature controlled cell, to a Faraday modulator and thence,
            through an analyzer onto a photomultiplier. The modulator is supplied with a crystal-controlled AC
            component. If an optically active sample is present,