            grating by means of a spherical mirror. The diffraction grating is movable, and so it can be set to
            monitor a particular wavelength that is characteristic of the element being measured, or it can be
            scanned to produce a complete absorption spectrum of the sample.




























                                                         Figure 2.19
                                            The Flame Atomic Absorption Spectrometer

            After leaving the grating, light of a selected wavelength, or range of wavelengths, is focused onto the
            photo cell. The position of the diffraction grating determines the wavelength of the light that is to be
            monitored. The flame absorption spectrometer is fairly sensitive, and can be readily used as a tandem
            instrument combined with a chromatograph, providing that an appropriate interface is employed.

            Chiro-Optical Measurements

            Over the last decade there has been a growing interest in chiral chromatography, a term  given to the
            separation of optically active compounds by chromatographic techniques. This interest has arisen
            largely as a result of the recognition that the majority of physiologically active substances exist in chiral
            form. Furthermore, in many countries, it is now mandatory that if a drug can exist with different chiral
            structures,