Page 288 - Tandem Techniques
P. 288

Page 272

            fluorescence. As many compounds do not fluoresce, the chromatogram provided by fluorescence
            monitoring is simpler than that by UV absorption, as only a fraction of the solutes are detected. This
            allows the peaks of interest to be highlighted against a relatively flat base line, and renders quantitative
            estimations far more accurate. This is clear from the separation depicted in Figure 7.13.



















                                                         Figure 7.13
                                        Comparison  of the Chromatograms of a Milk Analysis
                                           Monitored by UV Absorbance and Fluorescence
                                            Courtesy of the Hewlett Packard Corporation

            It is clearly seen that the area or height of the vitamin peak can be measured with far greater accuracy
            with the fluorescence chromatogram, as it is virtually devoid of interfering substances.

            The fluorescence spectrometer, used in conjunction with the liquid chromatograph, can often be
            programmed with respect to time. Thus, once the separation has been developed, the associated
            spectrometer can be programmed to provide the optimum excitation and fluorescence wavelengths for
            each peak as it emerges. This procedure provides the ultimate in sensitivity when using fluorescence
            detection. The principle of optimizing both the excitation and emission light wavelengths to obtain
            maximum sensitivity, however, can become quite complex as shown by the separation of some priority
            pollutants carried out on the PE LC/FL tandem instrument and depicted in Figure 7.14. The separation
            was
   283   284   285   286   287   288   289   290   291   292   293