            range the HP fluorescence spectrometer utilizes the xenon lamp. A diagram of the HP fluorescence
            spectrometer is shown in Figure 7.10.



































                                                         Figure 7.10
                                           The Hewlett-Packard Fluorescence Spectrometer
                                            Courtesy of the Hewlett Packard Corporation

            Light from the xenon source is focused onto an excitation grating which is movable to enable the
            wavelength of the excitation light to be selected. Light of the selected wavelength passes through a slit
            and is focused on the liquid chromatography flow cell. Light emitted at right angles to the excitation
            light is focused through a slit onto the emission grating and then through a third slit onto the
            photomultiplier. The emission grating is also movable, to enable the wavelength of the emission light to
            be selected or the fluorescence spectrum to be scanned. Between the second slit and the emission
            grating, there is a cut-off filter that prevents interference from stray light.