Page 290 - Tandem Techniques
P. 290

Page 274

            The gradient was linear and the flow rate was 1.3 ml/min. All the solutes are separated and the
            compounds, numbering from the left, are given in the Figure 7.14. The separation illustrates the clever
            use of wavelength programming to obtain the maximum sensitivity. The program is shown in Table 7.2.




















                                                          Table 7.2
                                                The Fluorescence Detector Program

            It is seen that during the development of the separation both the wavelength of the excitation light and
            that of the emission light was changed to an optimum for each particular solute. This ensured that each
            solute, as it was eluted, was excited at the most effective wavelength and then monitored at the
            strongest fluorescent wavelength.

            Monitoring the separation in this way is a somewhat elaborate procedure and must be carried out with a
            complex and relatively expensive instrument. Nevertheless, if the analysis is sufficiently important, this
            type of instrumentation may be essential. The tandem system can also provide fluorescence or
            excitation spectra, by arresting the flow of mobile phase, allowing the solute to reside in the detecting
            cell, and scanning the excitation and/or fluorescent light. This is the same technique as that used to
            provide UV spectra with the variable wavelength dispersion UV spectrometer. Thus it is possible to
            take excitation spectra at any chosen fluorescent wavelength, or fluorescent spectra at any chosen
            excitation wavelength. Consequently, many hundreds of spectra can be produced,
   285   286   287   288   289   290   291   292   293   294   295