Page 52

            tungsten filament lamp that can be made to either manually or automatically replace the deuterium
            lamp when the wavelength programmer has reached the appropriate transmission wavelength.

            The UV/Visible Dispersive Spectrometer


            A diagram of the optical system of a UV/Visible Dispersive Spectrometer is shown in Figure 2.3


































                                                          Figure  2.3
                                      The  Multiwave  Length  Dispersive  Spectrometer-Detector

            Light from the deuterium discharge lamp is collimated by two curved mirrors onto a holographic
            diffraction grating. The dispersed light is focused, by means of a curved mirror, onto a plane mirror and
            the light of the required specific wavelength selected by adjusting the angle of the plane mirror. Light of
            the selected wavelength is then focused by means of