Page 38 - A Practical Introduction to Optical Mineralogy
P. 38

THE  MICROSCOPIC STUDY OF MINERALS                                         POINTS  ON  USE  OF MICROSCOPE

                         that a load of 100 grams should be applied for 15 seconds. The size of    Table 1.2  Relation between  VHN and  Moh' s hardness.
                         the resulting square shaped impression depends on the hardness of the
                         mineral:                                                                          Moh's hardness (H)          - VHN
                                      VHN  =  1854  X  load
                                                 dz     kglmm 2
                                                                                                            1  talc                       10
                                                                                                            2  gypsum                     40
                         where  the  load  is  in  kilograms  and  d  is  the  average  length  of the      3  calcite                   100
                                                                                                            4  fluorite                  200
                         diagonals of the impression in  microns.
                                                                                                            5  apatite                   500
                           Hardness is expressed in units of pressure, that is, force per unit area.
                                                                                                            6  orthoclase                750
                         Thus the microindentation hardness of pyrite is  written:
                                                                                                            7  quartz                   1300
                                                                                                            8  topaz                    1700
                                    pyrite, VHN 100  =  1027-1240 kglmm 2                                   9  corundum                 2400
                                                                                                           [10  diamond]
                         The subscript 100 may be omitted as this is the standard load. As VHN
                         values are always given  in  kglmm this is  also often omitted.
                                                 2
                                                                                                     Thin sections must always be placed on the stage with the cover slip on
                          The determination of hardness is a relatively imprecise technique, so
                         an  average  of several  indentations  should  be  used.  Tables  of VHN   top  of  the  section,  otherwise  high  power  objectives  may  not  focus
                                                                                                   properly.
                         usually give a range in value for a mineral, taking into account variations
                         du~ to compositi~n.al variations, anistropy of hardness and uncertainty.    Polished samples must be level.  Blocks may be mounted on a small
                                                                                                   sphere of plasticine on a glass plate and pressed gently with a levelling
                         Br~ttleness,  plastiCity  and  elasticity  control  the  shape  of the  inden-
                         tatiOns,  and  as  the  shape  can  be  useful  in  identification  the  COM   device. Carefully machined polished blocks with parallel faces can usu-
                         r~commends that indentation shape (using the abbreviations given  in      ally  be  placed  directly  on  the  stage.  A  level  sample  should  appear
                         Fig.  1.10) be given  with  VHN values.                                   uniformly illuminated. A more exact test is to focus on the samples, then
                                                                                                   close the aperture diaphragm (seen using the Bertrand lens) and rotate
                          There is  a reasonable correlation between VHN and Moh's scratch
                         hardness as shown in  Table 1.2.                                          the stage. The small spot of light seen as the image should not wobble if
                                                                                                   the sample is  level.
                                                                                                     Good  polished  surfaces  require  careful  preparation  and are  easily
                        1.10  Points on the use of the microscope (transmitted and                 ruined. Never touch the polished surface or wipe it with anything other
                              reflected light)                                                     than  a  clean  soft tissue, preferably  moistened  with  alcohol  or a  pro-
                                                                                                   prietary cleaning fluid. Even a dry tissue can scratch some soft minerals.
                        Always focus using low power first. It is safer to start with the specimen   Specimens not in use should be kept covered or face down on a tissue.
                                                                                                     The  analyser  is  usually  fixed  in  orientation  on  transmitted-light
                        surface close to the objective and lower the stage or raise the tube to
                        achieve the position of focus.                                             microscopes but the polariser may be free to rotate. There is no need to
                                                                                                   rotate the polariser during normal use of the microscope and it should be
                                                                                                   positioned to give east- west vibrating polarised light. To check that the
                                                                                                   polars are exactly crossed examine an isotropic substance such as glass
                                                                                                   and  adjust  the  polariser  to  give  maximum  darkness  (complete
                                                                                                   extinction).
                                                                                                     The alignment of polariser and anlyser for reflected light can be set
                                   p            sf
                                (perfect)   (slightly fractured)   (fractured)                     approximately fairly easily. Begin by obtaining a level section of a bright
                                                                                                   i otropic  mineral  such as pyrite. Rotate the analyser and polariser to
                                                                                                   their zero positions, which should be marked on the microscope. Check
                                                                                                   that the  polars are crossed, i.e. the grain is  dark.  Rotate the analyser
                                                                                                   slightly to give as dark a field  as  possible. View the polarisation figure
            Figure 1.10                                                                            (see Section 1.4 ). Adjust the analyser (and/or polariser) until a perfectly
            Indentation                                                                            centred black cross is obtained. Examine an optically homogeneous area
                                         cc            cv
            shapes.                    (concave)     (convex)                                      of  a  uniaxial  mineral  such  as  ilmenite,  niccolite  or  hematite.  Using
                        26