STERILE MEDICAL DEVICE PACKAGE DEVELOPMENT  209

                                                     Operational Qualification (OQ)
                                    Investigate optimum   Visual inspection
                                    process parameters



                                                             Pkgs meet
                                                      NO    visual quality

                                                            YES

                                                         Packages produced @
                                                         optimum parameters



                                                          Visual inspection




                                                             Pkgs meet
                                                            visual quality
                                                      NO

                                                            YES
                                                             Leak Tests




                                                            Pkg integrity
                                                     FAIL
                                                           PASS

                                                         Production packages



                                                    Burst tests    Seal strength tests
                                               RESULTS                    RESULTS


                                                            Establish Pkg        GO TO
                                                            performance         PHASE 2
                                                            specifications

                                   FIGURE 7.2 Operational qualification flowchart example.