Page 135 - Chalcogenide Glasses for Infrared Optics
P. 135

112   Cha pte r  F o u r

              so because a bar or two were rejected when examined close to the
              beginning of the process. All chalcogenide glasses measured at AMI
              have had a rupture modulus greater than 2000 psi but less than
              3000 psi. The highest has been Amtir 1, as you would expect, at
              about 2700 psi. The value was confirmed twice in a Navy study. A
              different approach was taken at TI; plates were broken instead of
              glass bars. A set of metal concentric rings was used in place of bars
              in determining the rupture modulus.  Also, the smaller-diameter
              upper ring was connected using a swivel joint to ensure force was
              applied evenly to the surface of the plate. Figure 4.16 shows photo-
              graphs of a two 5.5-in-diameter plates of TI 1173 after being broken.
              One was 0.5 in thick and the other 0.25 in. The fracture patterns are
              clear in the photographs.

              4.3.4 Thermal Conductivity
              The absolute measurement of thermal conductivity for a solid is no
              easy task. Such a statement applies strongly for materials with low
              values. All glasses are considered poor thermal conductors because
              they are disordered solids. Chalcogenide glasses are even worse than
              oxide glasses. For many years, AMI has used the Thermal Comparator
              instrument developed by the Thermophysical Properties Research
              Center at Purdue University. A photograph of the instrument is seen
              in Fig. 4.14. The controls and microvoltmeter are the main unit while
              the thermal probe unit is on the right side. Also shown in the photo-
              graph is the acoustical thickness gauge unit to the left. The thermal
              probe has a thermal reservoir heated to 10 to 30°C above room tem-
              perature. The voltmeter is zeroed. A sample is placed over the hole in
              the sensing unit, and the lever is pushed down, which mechanically
              brings the probe in contact with the bottom of the sample and the
              Constantan tube leading from the heat reservoir. Heat begins to transfer
              from the reservoir to the sample. Chromel thermocouples on top of

















          Picture taken immediately after fracture,  With top plate removed,
              5.5 in diameter, 0.5 in thick   5.5 in diameter, 0.25 in thick
        FIGURE 4.16  Rupture modulus measurement of TI 1173 glass plates.
   130   131   132   133   134   135   136   137   138   139   140