Page 276 - Electrical Safety of Low Voltage Systems
P. 276

Applications of Electrical Safety     259





























                                  FIGURE 15.16 Isolating transformer supplying the circuits in the patient
                                  vicinity.


                                  patient vicinity, in fact, there may be small equipment, not supplied
                                  by the isolating transformer, requiring the ground connection avail-
                                  able at the local earthing bus. The above arrangement is a violation
                                  of the general rule, which prohibits the ECPs of separated systems
                                  to share the earth with nonseparated systems (prohibition also ap-
                                  plicable to Class II equipment). As seen, this rule intends to prevent
                                  enclosures of separated systems from becoming dangerously “live”
                                  due to potentials transferred by means of earthing connections.
                                     In the presence of the supplementary equipotential bonding in
                                  medical locations, this risk is, indeed, very low and deemed accept-
                                  able. In fact, even if the earthing bus attains a certain potential under
                                  fault conditions, all the ECPs in the patient vicinity will simultane-
                                  ously reach this same value, as Fig. 15.16 shows; ergo, no potential
                                  differences can appear among them and the patient is safe. As a con-
                                  sequence, the grounding connection in separated systems adopted in
                                  medical locations is deemed safe in the presence of the local equipo-
                                  tential bonding.

                                  15.8.4.1  Interruption of the Protective Conductor
                                            in Separated Systems
                                  The interruption of the PE is dangerous even in separated systems be-
                                  cause the resulting capacitive current through the patient may exceed
                                  the fibrillation limits (Fig. 15.17).
   271   272   273   274   275   276   277   278   279   280   281