Page 213 - Academic Press Encyclopedia of Physical Science and Technology 3rd Analytical Chemistry
P. 213

P1: ZCK/MBQ  P2: GQT Final Pages
 Encyclopedia of Physical Science and Technology  EN008B-382  June 30, 2001  18:58






               680                                                                                Liquid Chromatography























                      FIGURE 7 Diagram of a photodiode array instrument. For HPLC, the cuvette is replaced by a flow cell similar to that
                      used for the standard UV–VIS detector. [From Siouffi, A-M., Chapter 1, in “Food Analysis by HPLC.” (L. M. L. Nollet,
                      ed.), Marcel Dekker, New York.

               does not absorb, which is often the case, the reference side  which is a function of the solute concentration. Although
               is usually air. However, occasions arise where it is advan-  the RI detector is versatile, it is not particularly sensitive
               tageous to pump the mobile phase through both cells for  (microgram level) and is very prone to temperature fluc-
               subtraction purposes. The detector is quite stable to tem-  tuations. Careful temperature control of the detector cell
               perature fluctuations; however, air bubbles that lodge or  with a water jacket is crucial for maintaining a stable base-
               pass through the flow cell will cause baseline spikes. A  line. Modifications in the electronic design and the use of
               UV–VIS HPLC detector equipped with a monochromator  a laser source have also improved the detectability of the
               for specific wavelength selectivity is more versatile but  RI detector.
               also more expensive. Deuterium and tungsten lamps are  Luminescence detectors can be either based on fluo-
               used to provide ultraviolet and visible light, respectively.  rescence or chemiluminescence. The fluorescence HPLC
               Often multiple wavelengths can be monitored simultane-  detector is basically a fluorometer equipped with a flow
               ously. The photodiode array (PDA) UV–VIS detector has  cell slightly larger in volume than the UV–VIS detec-
               now become quite common as the standard HPLC de-  tor to permit more fluorescent light from the solute to
               tector ordered with a new instrument. The PDA detector  reach the photomultiplier. Again, the primary modifica-
               employs a reverse optics design which allows the entire  tion is the use of lenses to focus the excitation source light
               spectrum of light to be dispersed onto a diode array (see  onto the flow cell. Filter instruments are often more sensi-
               Fig. 7). Each diode of the array is responsible for detect-  tive than monochromator instruments because of greater
               ing light of a narrow wavelength range depending on the  light throughput. Again, lasers have allowed the use of
               desired spectral range to be covered and the speed of data  small volume flow cells with lower detection limits. The
               acquisition. This instrument can take a spectrum over a  fluorescence detector, although useful for only a select
               preselected wavelength range in a fraction of a second as  class of aromatic hydrocarbons or derivatives, is about
               the solute elutes from the column. Therefore, qualitative  100 times more sensitive than a UV–VIS detector. For
               (peak purity) as well as quantitative information about the  example, polyaromatic hydrocarbons can be determined
               unknown sample component can be obtained.         at the ppb level. Signal stability with respect to temper-
                 The differential refractometer is perhaps the second  ature is good. The coupling of fluorescence to HPLC
               most widely used HPLC detector because of its universal  has generally minimized background signals from im-
               nature. Potentially, any substance with a refractive index  purities and oxygen quenching effects. If the lamp of
               (RI) different from the mobile phase is detectable. The  the fluorescence detector is turned off, it can operate as
               deflection type RI detector consists of a light source pass-  an effective chemiluminescence detector. Postcolumn ad-
               ing two beams of monochromatic light through a double  dition of reagents such as luminol and a metal catalyst
               prism that constitutes the sample cell and reference cell.  for the detection of an oxidizing agent such as H 2 O 2 is
               If the mobile phase composition changes, the altered re-  necessary.
               fractive index causes the beam to be deflected from its  The electrochemical (EC) HPLC detector is basically
               initial position on the photomultiplier detector. The elec-  a small electrode composed usually of glassy carbon
               trical signal produced is proportional to the light position,  mounted in a flow cell (Fig. 8a). The auxiliary electrode is
   208   209   210   211   212   213   214   215   216   217   218