Page 236 - Fundamentals of Air Pollution 3E
P. 236

I. Analysis and Measurement of Gaseous Pollutants  199















                Fig. 14-2. Calibration apparatus for ozone analyzer (UV, ultraviolet).



        C. Nitrogen Dioxide
          The principal method used for measuring NO 2 is also based on chemilu-
        minescence (Fig. 14-3) (5). NO 2 concentrations are determined indirectly
        from the difference between the NO and NO X (NO + NO 2) concentrations
        in the atmosphere. These concentrations are determined by measuring the
        light emitted from the chemiluminescent reaction of NO with O 3 (similar
        to the reaction of O 3 with ethylene noted for the measurement of O 3),
        except that O 3 is supplied at a high constant concentration, and the light
        output is proportional to the concentration of NO present in the ambient
        air stream.
          Figure 14-4 illustrates the analytical technique based on this principle.
        To determine the NO 2 concentration, the NO and NO X (NO + NO 2) con-
        centrations are measured. The block diagram shows a dual pathway
        through the instrument, one to measure NO and the other to measure
        NO.,.. The NO pathway has an ambient air stream containing NO (as well
        as NO 2), an ozone stream from the ozone generator, a reaction chamber,
        a photomultiplier tube, and signal-processing circuitry. The NO X pathway
        has the same components, plus a converter for quantitatively reducing NO 2
        to NO. The instrument can also electronically subtract the NO from NO T
        and yield as output the resultant NO 2.













         Fig. 14-3. NO 2 chemiluminescent detection principle based on the reaction of NO
       with O 3.
   231   232   233   234   235   236   237   238   239   240   241