Page 112 - Fundamentals of Light Microscopy and Electronic Imaging
P. 112

OPTIMIZING THE MICROSCOPE IMAGE         95



                               field planes, and prepare a list of each set in order, beginning with the
                               lamp.
                             2. Adjust the condenser for bright-field mode. Image the diatom test plate
                               using green light (546 nm bandpass filter) and a 40  objective. Focusing
                               is difficult due to the extremely small size and transparent nature of the
                               specimen. First locate the diatoms under low power (10 ). Carefully
                               move the x and y coordinates of the stage until the specimen comes into
                               the field of view. Then swing in the 40  lens and refocus on the diatoms.
                               Compare the image of the diatoms with Figure 6-7 and note the indicated
                               spacings. Using bright-field optics and green light, open the condenser
                               aperture to the proper location using the telescope lens, and note the
                               species with the smallest spacing that it is possible to resolve with your
                               microscope. Indicate the species and note the distance between the striae
                               from the figure.
                             3. Calculate the theoretical resolution of the microscope under these condi-
                               tions. The NA is indicated on the barrel of the lens. Show your work. The
                               apparent and calculated resolution limits should roughly agree.
                             4. Now examine the diatom Pleurosigma with a dry 100  objective and
                               close down the field-stop diaphragm to illuminate just this one diatom and
                               no other. Make an accurate sketch of the hexagonal arrangement of the
                               pores. Close down the condenser diaphragm to its minimum size. Now
                               examine its diffraction pattern in the diffraction plane of the microscope
                               using the telescope lens. Make an accurate sketch of the diffraction pat-
                               tern.
                             5. How does the diffraction pattern relate to the spacing pattern of the
                               diatom pores in the image plane?
                             6. Examine the diatoms in red and blue light and with the condenser aperture
                               open or maximally closed. Which pair of conditions gives the lowest and
                               the best resolution of the striae?
                             7. Examine the diatom Pleurosigma or Diatoma with a 100  oil immersion
                               lens. Can the striae and pores be resolved? Examine the diffraction plane
                               under these conditions. Make sketches of both views. Now oil the con-
                               denser as well as the objective and repeat your observations. Make
                               sketches. Do you notice a difference in resolution?
   107   108   109   110   111   112   113   114   115   116   117