Page 155 - Fundamentals of Light Microscopy and Electronic Imaging
P. 155

138      POLARIZATION MICROSCOPY

                                this reason, manufacturers select out low strain objectives for use in polarization
                                microscopy.


                                ADJUSTING THE POLARIZING MICROSCOPE

                                 • Focus a specimen and adjust for Koehler illumination with both polarizing filters
                                    and compensator removed from the optical path.
                                 • Insert the fixed polarizer (depending on the microscope, either the polarizer or ana-
                                    lyzer might be fixed—i.e., glued into position in the holder so that it cannot rotate).
                                    If the polarizer is fixed, its transmission axis should be oriented horizontally east-
                                    west as seen looking in the microscope. If the polarizer is rotatable, check that the
                                    filter is positioned correctly by looking for the mark indicating the orientation of
                                    the transmission axis or checking the degrees of rotation on the holder.
                                 • Insert the rotatable analyzer (or polarizer), and rotate it until the two polars are
                                    crossed and maximum extinction is obtained. Do this by examining a blank region
                                    on a specimen slide and making fine adjustments to the analyzer until the image is
                                    maximally dark. The transmission axis of the analyzer should now be oriented ver-
                                    tically or north-south. The critical adjustment for extinction can only be made while
                                    looking in the microscope. Extinction indicates that the polars are crossed, but does
                                    not guarantee that their azimuths are perfectly oriented east-west and north-south.
                                 • Insert a telescope eyepiece or Bertrand lens and focus on the back aperture of the
                                    objective lens. A dark polarization cross is observed at extinction (Fig. 9-3), with
                                    brighter regions of intensity between the horizontal and vertical arms of the cross.
                                    Commonly, the horizontal and vertical components of the cross are broader in the
                                    center of the field and narrower at the periphery. This shape is caused by the depo-
                                    larizing effects of spherical lens surfaces in the condenser and the objective lenses,
                                    and is normal for most microscope systems. The cross should be centered in the
                                    field of view and should be perfectly upright. If not, the polarizer and analyzer may
                                    need to be rotated a bit to achieve proper orientation. Perfect alignment with respect
                                    to an azimuth at 0° is essential if quantitative measurements of azimuthal angles or
                                    birefringence are to be made with a compensator. Reference to the polarization

















                                                   (a)                            (b)
                                Figure 9-3
                                View of the polarization cross in the back aperture of the objective lens. Views before (a) and
                                after (b) proper adjustment of the condenser aperture diaphragm.
   150   151   152   153   154   155   156   157   158   159   160