Page 97 - Geochemistry of Oil Field Waters
P. 97

EMISSION SPECTROMETRY                                                 85


            hydrochloric  acid.  Evaporate  these  acid  solutions  to dryness,  dissolve the
            residues in distilled water, combine, and dilute to 1 liter with distilled water.
              Hydrochloric acid, concentrated.
              n-Propanol.



            Equipment.  The  necessary  equipment  includes  a  spectrograph;  a  d.c.  arc
            source, 18-A minimum; a plasma arc assembly; a plate-developing machine; a
            microphotometer;  50-ml volumetric  flasks;  10-ml microburets;  pipets;  and
            spectrographic plates, Eastman Kodak Type 1-N.

            Spectrochemical  excitation  conditions.  The  conditions  which  are  used  to
            determine barium, boron, iron, manganese, and strontium are as follows:
              Source, d.c. arc.
              Current, 18-25  A (keep constant).
              Voltage, 220 d.c.
              Pre-exposure time, 5 seconds.
              Exposure time, 15 seconds.
              Spectral region, 3200-5200  8, first order and second order.
              Dispersion,  reciprocal  linear  8.21 8/mm first  order,  4.02  8/mm second
            order or better.
              Plasma arc assembly.
              Helium lift gas, 7 liter per minute.
              Helium tangential gas, 60 liter per minute.
              Atomizer, Beckman Model 4030 with medium-bore capillary.
              Arc length above orifice, 7 mm.
              Full arc length, anode to cathode, 18 mm.
              Portion of  arc viewed, 2 mm above orifice.
              Orifice  electrodes; lower anode, Ultra  Carbon 106 drilled to 3.97  mm  in
           center hole, tapered to 9.53 mm at bottom; center ring, neutral Ultra Carbon
           861 drilled to 5.95 mm center hole, tapered to 9.53 mm at bottom.
              Cathode  electrode, vertical position,  Ultra  Carbon 6.35 mm  graphite rod
           with pointed tip.
              Slit, 20-p.
              Filter, 3-step.

           Spectrographic plate development conditions:  5 minutes in Eastman Kodak
           D-19 at 2OoC with  constant agitation;  30 seconds in stop bath at 3OoC with
           agitation; 5 minutes in Eastman Kodak rapid  fixer and hardener solution at
           2OoC with  constant agitation; 30 minutes in water  rinse with constant fresh
           supply  of  water;  30  seconds rinse  with  distilled  water;  and  30 minutes in
           constant air bath to dry.

           Microphotometer criteria. Slit, 12 p wide and 0.5 mm high.
              Read the background and intensity of  the following lines: Ba, 11, 4554.03
   92   93   94   95   96   97   98   99   100   101   102