Artificial Hearts                                            561






























              Fig. 34 Transcutaneous power leads provide power to the DeBakey axial heart pump.



                                                               Heart pump  M


                                        Skin                            Motor
                                                                        drive

               DC power             External   Internal   Power
                source    DC/AC       coil      coil      circuit     Controller


                          Controller
                                                               Aux power
                                                                 supply
              Fig. 35 Block diagram showing the components of a transcutaneous energy transfer
              system.



                 It uses rechargeable cells with the highest power density possible to min-
                 imize the mass that the patient needs to carry.
              •  DC/AC converter—this consists of an inverter or power oscillator
                 that typically generates an adjustable square-wave signal at a frequency
                 of between 100kHz and 1MHz.