Page 447 - Handbook of Materials Failure Analysis
P. 447

2 Experimental    445




                  the composition of sludge on the surface of the compressor wall from car air condi-
                  tioner are demonstrated. In addition, the identification of scaling of a valve block unit
                  from a failed medical oxygen resuscitation apparatus is presented.




                  2 EXPERIMENTAL
                  2.1 SAMPLES

                  Passenger car tires, car brake fluids, varnishes, rubber membrane from the pressure
                  vessel of an automotive integral accumulator, a part from car air conditioner, and a
                  valve block unit from the medical oxygen resuscitation device were used in the
                  investigation.
                     Poly(vinyl alcohol) (PVA) obtained from Celanese Chemicals Europe GmbH
                  (Frankfurt/Main, Germany) and polyvinylpyrrolidone (PVP) (Luvitec K 30) obtained
                  from BASF AG (Ludwigshafen, Germany) were used as standards for qualitative
                  identification of samples.



                  2.2 INSTRUMENTATION AND ANALYTICAL CONDITIONS
                  Approximately 100-200 μg of a solid sample were cut out with scalpel and inserted
                  without further preparation into the bore of the pyrolysis solids-injector and then
                  placed with a plunger on the quartz wool of the quartz tube of the furnace pyrolyzer
                  Pyrojector II™ (SGE Analytical Science, Melbourne, Australia) (Figure 17.2). The
                  pyrolyzer was operated at constant temperature of 550 or 700 °C. The pressure of
                  helium carrier gas at the inlet to the furnace was 95 kPa.
                     Py-GC/MS measurements were made by using two apparatuses. In the first
                  apparatus (1), the pyrolyzer was connected to a Trace 2000 GC (ThermoQuest/
                  CE Instruments, Milan, Italy) with a quadrupole MS Voyager (ThermoQuest/
                  Finnigan, MassLab Group, Manchester, UK) operated in electron impact ionization
                  (EI) mode. The fused silica GC capillary column (Elite-5ms 60 m long, 0.25 mm
                  I.D., 0.25 μm film thicknesses; PerkinElmer Instruments, Shelton, CT, USA) was
                  used. The gas chromatographic conditions were as follows: programmed tempera-
                  ture of the capillary column from 60 °C (7 min hold) at 5 °C min  1  to 100 °C and
                  then 10 °C min  1  to 280 °C (25 min hold at 280 °C). The temperature of the split/
                                                                       1
                                                                  3
                  splitless injector was 250 °C and the split flow was 10 cm min . Helium, grade
                  5.0 (Westfalen AG, Mu ¨nster, Germany) was used as a carrier gas at constant pres-
                  sure of 70 kPa during the whole analysis. The transfer line temperature was 280 °C.
                  The MS EI ion source temperature was kept at 250 °C. The ionization occurred
                  with a kinetic energy of the impacting electrons of 70 eV. The current emission
                  of the rhenium filament was 150 μA. The MS detector voltage was 350 V. Mass
                  spectra and reconstructed chromatograms (total ion current (TIC)) were obtained
                  by automatic scanning in the mass range m/z 35-450 u. Py-GC/MS data was
   442   443   444   445   446   447   448   449   450   451   452