Page 86 - Illustrated Pocket Dictionary of Chromatography
P. 86

82     FLASHBACK

            6
        is 10 , and the detection limits are in the parts per billion range.
        Because of the very low detection limits of this detector the input
        gases to the detector must be hydrocarbon free to keep the back-
        ground noise at an acceptable level. The FID has a linear range of
                       7
        approximately 10 .
        flashback    In GC systems using a vaporization inlet, flashback is the
        condition in which the sleeve volume is overloaded because of the
        large volume of expansion that occurs as a liquid sample is converted
        to a gas. This expansion causes the vapors to move back to the top
        of the inlet and recondense, leading to low and irreproducible results.
        Florisil A patented magnesium silicate (Mg 2 SiO 3 ) used predomi-
        nantly in sample cleanup.

        flow restrictor Most commonly used in SCF work and placed
        after a column to maintain constant system pressure during a
        separation.

        flow controller Used in GC systems to produce a constant gas
        flow in the system even though the resistance to flow may vary with
        time (e.g., as temperature changes during a gradient).

        flow injection analysis (FIA) The basic FIA technique is based
        on the injection of a sample into a nonsegmented continuous carrier
        stream that brings the analyte to the detector. Selectivity is either
        based on the purity of the sample (no other component in the system
        that the detector responds to) or a selective reaction (e.g., enzymatic
        or derivatization) of the analyte of interest. In more complex FIA
        systems there can be segmentation between samples. This is often
        accomplished through the injection of an air bubble between sample
        injections. Carry over and sample adsorption to the tubing can cause
        irreproducibility. Quantitation can be done either by peak height or
        peak area. Standards must be run along with samples.

        flow meter    Used to determine the volume (mL/min) flow rate of
        gases through the GC system. Two common flow meters are (1) soap
        bubble and (2) electrical sensor. The soap bubble flow meter is a
        manual operation in which a soap bubble moves up a calibrated
        volume tube and the time to move a given volume is recorded and
        from that the rate is calculated. The electrical sensor provides an
        automatic digital readout of the flow rate. Note that the flow rate is
   81   82   83   84   85   86   87   88   89   90   91