Page 540 - Instrumentation Reference Book 3E
P. 540

Detectors 523

            ters  are  used  in  radiocarbon  dating  systems
            because of their sensitivity for the detection of low-
            energj  I4C beta  particles  (E,,,,  = 156keV)  and
            even tritium 3H beta particles (E,,,,,  = 18 keV).


            22.2.1.1  Geiger-Mueller  detectors                   High voltage -
                                                       v
            The  Geiger  counter  has  been  and  is  the  most
            widely  used  detector  of  nuclear  radiation.  It   Figure 22.2  Geiger counter characteristic response
            exhibits several very attractive features, some of
            which are:                                sources.  Geiger  tube  manufacturers  normaily
                                                      quote the dead time of a particular counter, and
            (1)  Its  cheapness-manufacturing   techniques   it is customary to increase this time electronically
               have  so  improved  the  design  that  Geiger-
               Mueller  tubes  are  a  fraction  of  the  cost  of   in a unit often used with Geiger counters, so that
               solid-state or scintillation detectors.   the total dead  time is constant  but greater than
            (2)  The  output  signal  froin  a  Geiger-Mueller   any  variations  in  the  tube’s  dead  time,  since
                                                      between individual pulses the dead time can vary.
               tube can be of the order of  1 V, much higher
               than that from proportional,  scintillation, or   The counting rate characteristics can be under-
               solid-state detectors. This means that the cost   stood  by  reference to  Figure  22.2.  The  starting
                                                      voltage  V, is the voltage which, when applied to
               of the electronic system required is a fraction
               of that of other counters. A Geiger-Mueller   the  tube  viewing  a  fixed  radioactive  source,
               tube  with  a  simple high-voltage  supply  can   makes it just  start to count. As the high voltage
               drive most scaler units directly, with minimal   is  increased  the  counting  rate  rapidly  increases
               or no amplification.                   until  it reaches what  is called the plateau.  Here
            (3   The  discharge mechanism is  so  sensitive  that   the  increase in  counting  rate  from  VA to  VB  is
                                                      small, of  the  order of  1-5  percent  per  lOOV  of
               a single ionizing particle entering the sensitive
               volume of the counter can trigger the discharge.   high  voltage.  Above  VB  the  counting  rate  rises
                                                      rapidly  and the tube goes into a continuous dis-
               Nith  these  advantages  there  are,  however,   charge, which will damage the counter. An oper-
            some disadvantages which must be borne in mind.   ating point is selected ( VOp) on the plateau so that
            These include:                            any  slight variation  of  the  high-voltage  supply
                                                      has a minimal effect on the counting rate.
            (1)  The  inability  of  the  Geiger-Mueller  tube  to   In  order  to  count  low-energy  beta  or  alpha
               discriminate  between  the  energies  of  the   particles  with  a  Geiger  counter  a  thin  window
               ionizing particles triggering it.
            (2)  The tube has a finite life, though this has been   must  be  provided  to allow the particles to enter
                                                      the sensitive volume and trigger the counter. Thin-
               greatly extended bv the use of halogen fillings   walled glass counters can be produced, with wail
               instead of organic gases. The latter gave lives   thicknesses of the order of 30 mg/cm2, suitable for
               of  only about  10’” counts whereas the halo-   counting  high-energy beta  particles (in  this  con-
               gen tubes have lives of  lOI3 or more counts.
            (3)  Tlhere is a finite period between the initiation   text, the mass per unit area is more import$nt than
                                                      the  linear  thickness: for  glass, 25mg cm’  corie-
               of  a  discharge  in  a  Geiger-Mueller  counter   sponds to  about  0.1 mm).  For  low-energy betas
               and the  time when  it  will  accept  a new  dis-   or  alphas  a  very  thin  window  is called for, and
               charge. This is called the dead time, and is of   these have been  made with thicknesses as low as
               the order of 100 ,us.
                                                      1.5 mg  cm2. Figure  22.3  gives  the  transmission
              It is important to ensure that with Geiger count-   through windows of different thickness, and Table
            ers the  counting rate  is  such that  the dead-time   22.4 show how it is applied to typical sources.
            correction is only a small percentage of the count-   Alternatively,  the  source  can  be  introduced
            ing rate observed. This correction can be calcu-   directly into the counter by mixing as a gas with
            Iated from the relation:                  the counting gas, or if a solid source, by placing it
                                                      directly inside the counter and allowing the flow
                      R
                R’ = -                       (22.12)   of counting gas to pass through the counter con-
                    i  - RT                           tinuously. This is the flow-counter method.
            where  R  is  the  observed  counting  rate  per  unit
            time, R’  is the true counting rate per unit time and
            T is the counter dead time.               22.2.2  Scintillation detectors
              Dead  time  for  a  particular  counter  may  be   Scintillation  counters  comprise  three  main
            evaluated by a series of measurements using two   items: the scintillator, the light-to-electrical pulse
   535   536   537   538   539   540   541   542   543   544   545