Page 247 - Statistics for Environmental Engineers
P. 247

L1592_frame_C28.fm  Page 250  Tuesday, December 18, 2001  2:48 PM









                        It was necessary to measure the oxygen concentration in the influent to a pilot plant reactor. The influent
                       was under 20 psig pressure and was aerated with pure oxygen. The dissolved oxygen (DO) concentration
                       was expected to be about 40 mg/L. Sampling methods that are satisfactory at low DO levels (e.g., below
                       saturation) will not work in this situation. Also, conventional methods for measuring dissolved oxygen
                       are not designed to measure DO above about 20 mg/L. The sampling method that was developed involved
                       withdrawing the highly oxygenated stream into a volume of deoxygenated water, thereby diluting the DO
                       so it could be measured using conventional methods. The estimated in situ DO of the influent was the
                       measured DO multiplied by the dilution factor.
                        There  was a possibility that small  bubbles  would form and oxygen  would be lost as the pressure
                       dropped from 20 psig in the reactor to atmospheric pressure in the dilution bottle. It was essential to
                       mix the pressurized solution with the dilution water in a way that would eliminate, or at least minimize,
                       this loss. One possible technique would be to try to capture the oxygen before bubbles formed or escaped
                       by introducing the sample at a high rate into a stirred bottle containing a large amount of dilution water.
                       On the other hand, the technique would be more convenient if stirring could be eliminated, if a low
                       sample flow rate could be used, and if only a small amount of dilution water was needed. Perhaps one
                       or all of these simplifications could be made. An experiment was needed that would indicate which of
                       these variables were important in a particular context. The outcome of this experiment should indicate
                       how the sampling technique could be simplified without loss of accuracy.
                        Four variables in the sampling procedure seemed critical: (1) stirring rate S, (2) dilution ratio D, (3)
                       specimen input location L, and (4) sample flow rate F. A two-level, four-variable fractional factorial
                             4−1
                       design (2 ) was used to evaluate the importance of the four variables. This design required measurements
                       at eight combinations of the independent variables. The high and low settings of the independent variables
                       are shown in Table 28.1. The experiment was conducted according to the design matrix in Table 28.2,
                       where the factors (variables) S, D, L, and F are identified as 1, 2, 3, and 4, respectively. The run order
                       was randomized, and each test condition was run in duplicate. The average and difference between
                       duplicates for each run are shown in Table 28.2.



                                     TABLE 28.1
                                     Experimental Settings for the Independent Variables
                                                 Stirring   Dilution   Sample Input   Sample Flow
                                     Setting       S     Ratio D   Location L   Rate F
                                     Low level (−)  Off    2:1      Surface    2.6 mL/sec
                                     High level (+)  On    4:1      Bottom     8.2 mL/sec





                            TABLE 28.2
                            Experimental Design and Measured Dissolved Oxygen Concentrations
                                                        Duplicates (mg/L)  Avg. DO (mg/L)  Difference (mg/L)
                            Run  S (1)  D (2)  L (3)  F (4)  y 1i  y 2i     y i           d i
                             1    −     −     −    −     38.9    41.5      40.20         −2.6
                             2    +     −     −    +     45.7    45.4      45.55          0.3
                             3    −     +     −    +     47.8    48.8      48.30         −1.0
                             4    +     +     −    −     45.8    43.8      44.80          2.0
                             5    −     −     +    +     45.2    47.6      46.40         −2.4
                             6    +     −     +    −     46.9    48.3      47.60         −1.4
                             7    −     +     +    −     41.0    45.8      43.40         −4.8
                             8    +     +     +    +     53.5    52.4      52.95          1.1
                            Note: Defining relation: I = 1234.

                       © 2002 By CRC Press LLC
   242   243   244   245   246   247   248   249   250   251   252