Page 160 - Tandem Techniques
P. 160

Page 143

            an appropriate solvent. Alternatively, it can be trapped on a suitable adsorbent contained in a packed
            tube, and then thermally desorbed into the carrier gas of a gas chromatograph.




























                                                          Figure 4.9
                                              The Separation of a Sample of Basil Oil
                                             Courtesy of the Perkin Elmer Corporation

            The column that was used in this example, was a macro-bore open tubular column, 50 m long, 0.32 mm
            in diameter and carried a 5 mm film of a methyl silicone. The chromatogram was obtained by plotting
            the integrals of each adsorption curve against time. The chromatogram of oil of basil, obtained in this
            way, is shown in Figure 4.9. The light pipe was maintained at 250°C and the scavenger flow was set at
            1 ml/min. Each point on the chromatogram resulted from the sum of two consecutive spectra. A
            spectrum, taken at the peak maximum of linalool (peak 4), is shown in Figure 4. 10. It is seen that a
            clean spectrum is obtained and the resolution from the macro-bore capillary column does not appear to
            have been denigrated. It has been claimed that spectra, with adequate resolution and signal-to-noise for
            solute identification, have been obtained from as little as 10 ng of material. However, the minimum
            sample size that can
   155   156   157   158   159   160   161   162   163   164   165