            cage of solid argon and, as the argon insulates the sample from the gold surface, there can be no
            catalyzed decomposition, or molecular rearrangement. The carrier gas is removed by a low-pressure
            turbo-molecular pump that contains no oil, and thus eliminates the possibility of any long-term sample
            contamination.





































                                                         Figure 4.14
                                             The Mattson GC/IR Cryostatic  TM   Interface
                                               Courtesy of Mattson Instruments Inc.

            Due to the extremely inert condition of the trapping system, the samples can be held for a very long
            time on the drum without loss or change. Consequently, the spectrum  of any particular sample can be
            taken repeatedly, any number of times, to improve the signal-to-noise ratio and thus the overall
            sensitivity.