234 DIAGNOSTIC EQUIPMENT DESIGN
B
B Gz y G2
y
z
x
B G1
φ
z/a
FIGURE 8.3 A filamentary, single turn, saddle transverse gradient coil set is illustrated.
In addition, patterns for a typical unshieled transverse gradient coil is shown.
8.3.3 Acoustic Noise
Time-varying magnetic field gradients spatially encode the anatomy of the patient in MR imaging.
Switched gradients may also produce acoustic noise. There are two sources of gradient-generated
l
acoustic noise. A conductor of length carrying current I in a static magnetic field B 0 will experi-
ence a force F m due to the magnet 48,49 :
F=Il × B 0 (8.13)
m