Page 260 - Biomedical Engineering and Design Handbook Volume 2, Applications
P. 260

DESIGN OF MAGNETIC RESONANCE SYSTEMS  239

                            At higher field strengths, most of the RF losses are due to patients. However, for low field
                          systems, patient losses are much smaller than coil losses. One approach for reducing the SNR
                          impact of such low field coil losses is to use superconducting surface coils. 94  These coils are
                          typically limited in size and require attached cryostats. In addition, very high Q translates to
                          very narrow bandwidths and tight tolerances on tuning.

                          Receiver Coil Arrays.  Signal-to-noise ratio may be improved by replacing a single, large field of
                          view coil with an array of smaller coils that combine for a field of view similar to the larger coil.
                          Each smaller coil receives less noise than the larger coil. Since noise is not correlated, while signals
                          are (due to coil design) signal to noise is typically higher when receiver coil arrays are used.
                          Typically, noise from adjacent coil elements is designed to be uncorrelated by appropriate overlap or
                          circuit design.

                          Parallel Imaging.  Magnetic resonance parallel imaging techniques use multiple receiver coils
                          (with different spatial sensitivities). These receiver coils are used to supply some portion of the spatial
                          encoding information to reduce the total number of gradient phase encodings. 95–101  The techniques
                          typically are used to obtain shorter scan times and may shorten readout times. Some image artifacts
                          can be reduced using parallel imaging.
                            Typically, in parallel imaging receiver coil spatial sensitivity is determined from low-resolution
                          reference images. Then aliased data sets can be generated for each individual coil. Receiver coil
                          spatial sensitivity information may then be used with either image-domain or frequency (k-space)-
                          domain reconstruction techniques to generate full field of view (nonaliased) images.
                            Signal to noise in parallel imaging is reduced by the shorter imaging time and by nonideal coil
                          geometry. Nonideal coil geometry finds expression in the g factor. The g factor refers to the pixel-
                          by-pixel signal-to-noise ratio obtained with parallel imaging divided by the signal-to-noise ratio
                          using equivalent conventional techniques with the same receive coils. The g factor is only defined in
                          regions where signal to noise is above a threshold and technically applies only to image-based par-
                          allel imaging methods.
                            The reduction of the number of phase-codings needed due to multiple receiver coils in parallel
                          imaging is called the acceleration factor. Maximum acceleration factors are somewhat less than the
                          number of parallel imaging receiver coils.



              8.4.4 Preamplifiers
                          Low-noise figure preamplifiers with noise impedances matched and relatively near the receiver
                          coils are required to avoid degrading SNR. Quadrature systems with preamplifiers on each channel
                          have small SNR advantages over quadrature systems employing low loss combiners and a single
                          preamplifier.


              8.4.5 SAR

                          During MR scans below 3 T or so, RF power deposition in patients can be approximated from
                          quasistatic analysis, assuming electric field coupling to patients can be neglected. 10,34  Let R be the
                          radius, σ the conductivity, and ρ the density of a homogeneous, sphere of tissue (Fig. 8.1). Assume
                          that this sphere is placed in a uniform RF magnetic field of strength B and frequency ω. Let
                                                                                1
                          the radiofrequency duty cycle be η. Then average specific absorption rate (SAR), SAR , may be
                                                                                          ave
                          expressed as 63
                                                             σηω 2  2 1 BR  2
                                                      SAR ave =                              (8.20)
                                                                20ρ
   255   256   257   258   259   260   261   262   263   264   265