IEC 60601-2-33
                                                  IEC 60601-2-33
                                                    controlled mode
                                                       IEC 60601-2-33 second controlled mode (operator takes deliberate  (needs IRB approval) action to enter) normal mode  4 T  >  B 0 4 T     B 0 2 T <  4 W/kg Whole-body ave SAR >  4 W/kg Whole-body ave. SAR    2 W/kg  with 6 min ave. with 6 min ave.  3.2 W/kg Ave. head SAR >  3.2 W/kg  Ave. head SAR    3.2 W/kg   with 6 min ave. with 6 min ave.  10 W/kg in head Local SAR >  10 W/kg Local SAR   10 W/kg in head  20 W/kg in extremities or trunk, >  20 W/kg in head or trunk,




                                                             2.0 T  Whole-body ave. SAR    with 6 min ave.  Ave. head SAR    with 6 min ave.  Local SAR    or trunk,    with 6 min ave.  6 min ave.  long-term SAR level     nerve mean

                                                                                  SAR
                                                             B 0                                    dB/dt
                                                                4 W/kg


                                                  FDA significant risk   criteria (http://www.  fda.gov/cdrh/ode/  magdev.html)  Whole-body ave. SAR ≥  3 W/kg   8 W/kg in any   gram (head or torso)  Sufficient for severe discomfort



                                             MR Safety Standards  4 T  >  B 0  with 15 min ave.  Ave. head SAR >  with 10 min ave.  Local SAR >  N/A  N/A











                                             TABLE 8.1    Safety parameter  B 0  WB SAR  Head SAR  Local SAR  Partial body SAR  Short-term SAR  dB/dt