272 10 The Use of SMB for the Manufacture of Enantiopure Drug Substances: From …
The total column length is:
L = 1.43 m (8 columns of 18 cm length)
The velocity is then:
u = 0.0064 m s –1
The column diameter is then determined using the mean flow rate in the SMB.
∅= ⋅ 4 Q mean = 049 m
.
π ⋅u
and Equation (21) gives the shift period: ∆T = 0.92 min.
Using simulation software (essentially solving the mass balance equation in the
transitory regime), one can show that the internal profiles of the products in the sys-
tem are given by the profiles presented in Figure 10.6. Samples were taken between
columns at half periods.
C (g/L)
0,6
0,5
0,4
0,3
0,2
0,1
0
Eluent Extract Feed Raffinate
Figure 10.6. Internal profile (simulated) obtained for the separation under linear conditions.
This simulation shows that the purity of the raffinate collected over a complete
cycle is about 97.0 % and the extract is about 98.5 %. These results are close to the
target and adequate purity can be reached with minor changes in flowrates. Further
details will be given in the next section.