308 Chapter 13





























                                                Figure 13.22
             Step 4—The release of tension in the mechanically driven nylon thread attached to the flexible tip
            via rotational stepper motors will allow the nasopharyngoscope to revert to its straight formation.
                 The device can then be removed from the nasopharyngeal cavity via the reversal of the
                                          translational stepper motor.


            13.6 Design verification

            Design verification is an essential step in the development of any prototype, as it ensures
            that the prototype was fabricated as intended (according to the design acceptance criteria).


            13.6.1 Design parameter verification

            Before verification for design parameters, we perform kinematic and mathematical analysis
            of BEE in order to determine theoretical values by which we could gauge the performance
            of the prototype. The calculations and relevant illustrations are shown as follows
            (Figs. 13.23 and 13.24):
            The theoretical maximum bending angle at joint:


                                                           h 0
                                            θ max 5 2 arctan  :
                                                           D
            Substituting the following values:
                                       h 0 5 2:6 mm and D 5 6:0mm;