224             15. Air Pollution Monitoring and Surveillance

       Table 15-2 contains a specific example of these components for ambient
       monitoring for ozone.
         In addition to fulfilling the in-house requirements for quality control, state
       and local air monitoring networks which are collecting data for compliance
       purposes are required to have an external performance audit on an annual
       basis. Under this program, an independent organization supplies externally
       calibrated sources of air pollutant gases to be measured by the instrumenta-
       tion undergoing audit. An audit report summarizes the performance of the
       instruments. If necessary, further action must be taken to eliminate any
       major discrepancies between the internal and external calibration results.
         Data quality assessment requirements are related to precision and accu-
       racy. Precision control limits are established, i.e., +10% of span value, as
       calculated from Eq. (15-1). The actual results of the may be used to calculate
       an average deviation (Eq. 15-3):



       where d t is the percentage difference, y, the analyzer's indicated concentra-
       tion of the test gas for the zth precision check, and x f the known concentra-
       tion of the test gas for the fth precision check.







                                    TABLE 15-2
                  Quality Control Components for Ambient Ozone Monitoring
            Component                         Description

        Method             Chemiluminescent O 3 monitor
                           Calibration method by certified ozone UV transfer method
       Manifold/instrument  Instrument connected to sampling manifold which draws ambient
         configuration      air at 3 m into instrument shelter
       Calibration         Multipoint calibration on 0.5-ppm scale at 0.0, 0.1, 0.2, and
                            0.4 ppm weekly
       Zero/ span check    Zero check ± 0.005 ppm
                           Span check 0.08-0.10 on a 1.0-ppm full scale daily
       Control specification  Ethylene flow
         checks            Sample flow, daily
       Corrective limits   ± 0.005 ppm zero and span
       Corrective action   Do multipoint calibration; invalidate data collection since last zero/
                            span check within control limits
       Preventive maintenance Manufacturer's procedures to be followed
       Recording and       Data reported weekly to quality assurance coordinator, with invalid
         validating data    data flagged
       Documentation       Data volume includes all quality control forms, e.g., zero/span
                            control charts and multipoint calibration results