Page 215 - Fundamentals of Light Microscopy and Electronic Imaging
P. 215

198      FLUORESCENCE MICROSCOPY

                                separate filter sets to capture different images of each of the two fluorochromes. Bleed-
                                through can be reduced, but is never completely eliminated, because:

                                 • The excitation spectra of two or more fluorochromes are broad and overlap to a sig-
                                    nificant extent (Figs. 11-5 and 11-6; Color Plate 11-1). Thus, the excitation of fluo-
                                    rescein at 490 nm also causes rhodamine to fluoresce. The solution is to choose
                                    fluorochromes with well-separated excitation spectra so that the excitation peak for
                                    one fluorochrome is many wavelengths shorter than the excitation peak for the sec-
                                    ond fluorochrome.
                                 • The fluorescence spectra of the fluorochromes may overlap, allowing fluorescence
                                    from the shorter-wavelength fluorochrome to contribute to the image of the longer-
                                    wavelength dye. To minimize the problem, a narrow bandpass filter is used as an
                                    emission filter to collect only the peak fluorescence of the lower-wavelength fluo-
                                    rochrome; a long-pass barrier filter is usually used for the longer-wavelength fluo-
                                    rochrome.
                                 • The fluorescence emission of one fluorochrome (fluorescein) may stimulate a sec-
                                    ond longer-wavelength dye (rhodamine) to fluoresce. Selecting well-separated dyes
                                    and assuring that labeling and fluorescence of the dyes are balanced help reduce
                                    this problem.
                                 • If the amount of labeling and the intensity of fluorescence of the two fluorochromes
                                    are not equally balanced, the brighter signal can overwhelm and penetrate the filter
                                    set for the second signal and cause a significant contribution to the image of the sec-
                                    ond dimmer signal. The intensity of fluorescence from dyes such as fluorescein and
                                    rhodamine should be similar and is adjusted according to the amount of dye in the
                                    specimen and the type of illumination used. Microscopists using epi-illumination
                                    with a mercury arc lamp frequently forget that rhodamine is excited 10  more
                                    effectively than fluorescein owing to the bright 546 nm emission line in the mercury
                                    arc spectrum.

                                    Even when these factors are controlled, the amount of signal crossover and bleed-
                                through generally remains about 10–15%. In experiments involving double staining, you
                                should always examine single-stained specimens using the filter set for the other fluo-
                                rochrome to assure that the amount of bleed-through is minimal. In many cases, the main
                                cause of bleed-through is unequal staining by the two dyes. Remember that it is nearly
                                always desirable to return to the lab bench to prepare a proper specimen, rather than to use
                                unequal exposures and image processing on a poorly prepared specimen just to save time.


                                EXAMINING FLUORESCENT MOLECULES IN LIVING CELLS


                                Fluorescence microscopy is potentially damaging to living cells, since light sources are
                                intense and rich in damaging UV and IR wavelengths, and because fluorescence filter
                                sets are not totally efficient in removing these unwanted wavelengths. The chief concern
                                is phototoxicity to the cell through absorption of photons by introduced and endogenous
                                fluorophores and the generation of reactive oxygen species, including singlet oxygen
                                 1

                                ( O ), superoxide (O • ), hydroxyl radical (OH•), various peroxides (ROOR ),
                                                  2
                                   2
                                hydroperoxides (ROOH), and others. They react with oxidizable metabolites and com-
                                ponents in the cell such as the pyridine nucleotides in nucleic acids, several amino acids,
                                glutathione, lipids, and ascorbate. Among the immediate effects is damage to membrane
   210   211   212   213   214   215   216   217   218   219   220