Page 186 - Gas Purification 5E
P. 186

-4lkanolaniines for Hydrogen Surfide and Carbon Dioxide Removal   173

                 (equivalent to 2,500 gaY(day)(ft2) used to set the settler dimensions. Separators downstream
                 of contactors, using either random packing or sieve trays, can be smaller than those follow-
                 ing mixers or eductors because the degree of agitation and, therefore, the formation of fine,
                 entrainable droplets is less in the packed or tray columns. Stewart and Laming (1991,
                 1994A, Bj recommend LPG residence times of  15 minutes for a gravity settler downstream
                 of a pzcked bed LPG treater.

                 Amine Filters
                   Figure 2-96 shows an optional amine filter on the lean amine solution flowing to the
                 LPG/amine contactor. Although many LPG maters do not have  dedicated mechanical or
                 activated carbon filters, their use should be considered. Bacon (1972) recommends full flow
                 mechanical and activated carbon lean amine filters for LPG treaters. Russell (1980) reports
                 particulate plugging of the random packing beds used in LPG treaters and recommends full
                 flow amine filtration to minimize plugging. DuPart and Marchant (1989) also recommend
                 both full flow mechanical and activated carbon filtration of  the amine flowing to LPG
                 treaters. As noted by both DuPart and Marchant (1989) and Strigle (1994), failure to remove
                 impuritie.s can reduce the capacity of the LPG treating system.
                 Removal of COS from LPG by Amines

                   A common specification for LPG is the Copper Strip corrosion test. This test indicates the
                 level of elemental sulfur and H2S by  the degree of  darkening of  a strip of  pure copper. A
                 #1A Copper Strip test result corresponds to 1-2  ppmw or less H2S in the LPG. COS in dry
                 LPG is not detectable by the Copper Strip test (Bacon, 1972; Perry,  1977A, Bj. However, if
                 the LPG comes in contact with water, e.g.,  during shipment in a tank car containing water,
                 the COS hydrolyzes to H2S and C02 and the LPG can fail the Copper Strip test upon retest-
                 ing (Holmes et al., 1984: Perry, 1977A, B). The COS hydrolysis reaction is
                   COS + H20 = H2S + CO?                                        (2-49)


                   Because of  this possible reaction, COS removal during amine LPG treating is desired.
                 DGA, MEA, and possibly DIPA can react with COS during the amine treating operation so
                 that a #lA Copper Strip test result can be achieved even if the LPG is wet. DGA is reported
                 to reduce COS levels from 4,500 ppmw to less than 2 ppmw (McClure and Morrow., 1979A1
                 B, Cj. The DGA reacts with COS to form a degradation product that can be converted back
                 to DGA by thermal reclaiming. MEA reacts hversibly with COS and, when MEA is used
                 for LPG treating, the treated LPG can pass the #1A Copper Strip test. However, MEA is not
                 preferred for simultaneous C02,  H2S, and COS removal from LPG because of the amine loss
                 caused by the irreversible reaction with COS (McClure and Morrow, 1979C). DIPA is also
                 reported to be capable of removing COS to several ppmw  (S.I.P.M.,  1979). Mick (1976)
                 reports that DEA can only remove COS to 90-135  ppmw. Bacon and Pearce (1985) report
                 that MDEA can remove only 33%.  of the COS in LPG and cannot remove COS to levels low
                 enough to pass the #1A Copper Strip test.
                   Although neither MDEA nor DEA can remove COS to levels low enough to pass the Cop-
                 per Strip test, use of  a formulated caustic wash downstream of  the amine/LPG treater is
                 reported to remove COS to levels significantly less than 1 ppmw (Bacon and Pearce, 19853.
                 The composition of the formulated caustic may be based on technology outlined by Johnson
   181   182   183   184   185   186   187   188   189   190   191