Page 160 - Glucose Monitoring Devices
P. 160

Clinical accuracy   161




                               400




                               300

                              CGM BG (mg/dl)   200








                               100




                                0
                                 0        100       200        300      400
                                              Reference BG (mg/dl)
                  FIGURE 8.1
                  An ensemble of CGM data points plotted over the original error grid. Each CGM data point
                  is represented by a vector that has a position and direction of change.
                  analysis, known as the Clarke EGA [49]da two-dimensional plot of CGM BG
                  levels (y-axis) plotted against reference BG measured at the same time (x-axis):
                     More information means a better chance of making a correct decision, but also
                  means that accuracy must be assessed on more than one component. What is needed
                  is an error grid that quantifies the clinical accuracy of CGM by both absolute BG
                  value (point accuracy) and change in BG (rate accuracy), and takes into account
                  the biological factor of the time lag between blood and interstitial tissue. This
                  continuous glucose-error grid analysis (CG-EGA) also needs to be applicable to
                  the original EGA [49], the consensus EGA (CEG, [50]), and the surveillance error
                  grid (SEG, [51]). In addition, it is imperative that the distribution of hypoglycemic,
                  euglycemic, and hyperglycemic BG levels evaluated be similar to that observed in
                  individuals with diabetes because rates and direction of BG change may have
                  different clinical interpretations in different ranges. For example, a low BG that is
                  falling rapidly signals a different treatment decision than a low BG level that is rising
                  rapidly. An error in rate or direction in the low BG change could lead to a very
                  dangerous clinical decision.
                     The CG-EGA includes two different grids: one for point accuracy (P-EGA) and
                  another for rate accuracy (R-EGA). When the reference rate of BG change is
                  within  1 to 1 mg/dL/min, it is considered to not be clinically significant and the
                  original EGA is used. Thus, in this case, the P-EGA is equivalent to the original
                  EGA [49] depicted in Fig. 8.2 and represents a scatterplot of CGM versus Reference
   155   156   157   158   159   160   161   162   163   164   165