Page 162 - Glucose Monitoring Devices
P. 162

Clinical accuracy   163




                          1: Reference Rate of [-2,-1) mg/dl/min:  3. Reference Rate of (1,2] mg/dl/min:
                       400                              400
                                 uCp                    300  uEp  uCp
                   Moderate BG Fluctuation  Estimated BG (mg/dl)  200  uDp  uBp  lBp  Ap  Estimated BG (mg/dl)  200  uDp  uBp  lBp  Ap
                       300
                           uEp




                       100

                                            lEp lDp     100                  lEp lDp
                        0          lCp                   0          lC
                          0    100   200    300   400     0     100   200    300  400
                               Reference BG (mg/dl)             Reference BG (mg/dl)
                           2. BG falls faster than –2 mg/dl/min:  4. BG rises faster than 2 mg/dl/min:
                       400                              400
                                 uCp                    300  uEp  uCp
                       300
                   Rapid BG Fluctuation  Estimated BG (mg/dl)  200  uDp  uBp  lBp  Ap  Estimated BG (mg/dl)  200  uDp  uBp  Ap
                           uEp




                       100

                                            lEp lDp     100            lBp   lEp lDp
                        0          lCp                   0          lC
                          0    100   200    300   400     0     100   200    300  400
                               Reference BG (mg/dl)             Reference BG (mg/dl)
                  FIGURE 8.3
                  Expansion of the P-EGA zones compensating for interstitial time lag at different rates of
                  blood glucose change. Upper zones are expanded to compensate for falling BG levels;
                  lower zones are expanded to compensate for increasing BG levels.

                     The rate error grid (R-EGA) is derived by plotting CGM BG rates of change
                  versus reference BG rates of change (Fig. 8.4), For this analysis, it is suggested
                  that a sampling frequency of at least 10e15 min be used. This frequency should
                  be sufficient to capture representative BG changes. Sampling should include
                  blocks of time (w4 h) representative of changes in subjects’ conditions (food,
                  exercise, insulin administration). The boundaries of the R-EGA have been set
                  at  4to þ4 mg/dL/min based on maximum and minimum BG fluctuations
                  collected from numerous datasets. The R-EGA is divided into zones A through
                  E, which have clinical treatment designations similar to the P-EGA:
                  1. In zone A, the CGM rate of change is within  1to þ1 mg/dL/min of reference
                     rate and is therefore clinically accurate. The zone A boundaries are expanded
                     to  2 and þ2 mg/dL/min at extreme rates of  4 and þ4 mg/dL/min because
                     these rates of rise and fall are rare, and correctly recognizing their direction
                     results in a clinically accurate decision;
   157   158   159   160   161   162   163   164   165   166   167