3 Results    467


















                  FIGURE 17.17
                  Residue on the surface of a valve block unit from the failed medical oxygen resuscitation
                  device.



































                  FIGURE 17.18
                  Pyrolysis-GC/MS chromatogram at 700 °C of the residue on the surface of the valve block unit
                  from the failed medical oxygen resuscitation device. Pyrolysis-GC/MS apparatus 2,
                  chromatographic conditions 2. Peak identification: 1—tetrafluoroethylene; 2—benzene;
                  3—hexamethylcyclotrisiloxane; 4—not identified; 5—styrene;
                  6—octamethylcyclotetrasiloxane; 7—oxazepam; 8—triazolam; 9—not identified;
                  10—phenytoin; 11—clozapine; 12—nordazepam; 13—levomepromazine; 14—not
                  identified; 15—clomipramine; 16—buprenorphine.