Page 288 - High Temperature Solid Oxide Fuel Cells Fundamentals, Design and Applications
P. 288

Testing of  Electrodes, CelIs and Short Stacks  265

           Note that there is a potential loss across the electrolyte due to the electrolyte
           resistance. The potential steps at the interfaces are now smaller compared to the
           OCV condition due to the losses originating from the polarisation resistance of
           the electrode processes. Figure 10.2~ illustrates the case of an anode-supported
           cell. The potential is given for both the position of the ‘reference’ electrode where
           no current flows across the electrolyte (i.e. the electrical potential is constant
           across the electrolyte), and for a position far away from the electrode edges as in
           Figure 10.2b.
             As the potential along the anode, which is a very good electronic conductor, is
           the same everywhere (the anode constitutes an isopotential plane), the potential
           of the anode at the ‘reference’ electrode must be equal to the potential in the
           middle of the current-bearing part of the anode. Thus the two potentia1 curves in
           Figure 10.2~ must start at the same point. Therefore, as seen in Figure 10.2c, the
           potential difference between the ‘reference’ electrode and the upper electrode in
           Figure 10.1 is simply the total polarisation of the full cell. Thus it is clear that the
           ‘reference’ electrode measures  only  the  emf  of  the  cell with  the actual  gas
           compositions at the  ‘reference’ and inside the support at the lateral position
           opposite to the reference while the current is flowing. Thus, no information on
           what happens  on  any of  the  working  electrodes can be  derived from  such
           measurements. If the concentrations of the reactants and products at the lateral
           position of  the reference  electrode were  the same as in the active electrode/
           electrolyte interfaces, then it would be possible to deduce the total concentration
           overpotential.  This is, however, in general not the case, and this means that
           the  voltage  difference  actually  measured  between  the  working  and  the
           ‘reference’ electrodes cannot be assigned  any clear  meaning.  In the present
           context, it is also helpful to remember that a single-electrode potential cannot be
           measured directly: it is only possible to measure a potential difference between
           two electrodes.
             To avoid such problems, it is necessary to test the electrodes using a suitable
           three-electrode  set-up  or a  symmetrical two-electrode cell [40], even though
           both  have  their  shortcomings. Some examples of  useful  set-ups for studying
           electrode performance are briefly presented here. The set-ups are based on zirconia
           pellets  with  an  eIectrode  arrangement  suitable  for  three-electrode  studies.
           Specific material choices and mounting details are given in the caption to Figure
           10.3. Such pellet-like geometries, where the reference electrode can be suitably
           placed (in a bore as in the figure or as a ring around the pellet) are suitable for
           fundamental studies  of  electrode kinetics.  The pellet-like  test  cell geometries
           depicted in Figure 10.3 suffer from two disadvantages: it is difficult to ensure that
           the fabrication process for the electrodes used is identical to the one used for the
           actual cells, and the ohmic resistance between the working and the reference
           electrodes is quite substantial which may result in a ‘signal to noise’ problem
           when very good electrodes are studied. An improved three-electrode geometry
           using a ring-shaped working electrode is currently being investigated [3 91.
             To measure a particular  electrode performance in detail, a symmetrical cell
           with identical electrodes on each side can be used as shown in Figure 10.4. This
           has a platinum mesh to make good contact with the electrodes and two platinum
   283   284   285   286   287   288   289   290   291   292   293