Page 177 - Instrumentation Reference Book 3E
P. 177

Analysis methods that measure size directly  161
             of light. At these very small sizes, however. inter-
             ference and  diffraction  effects  cause  significant
             errors,  and  below  the  wavelength  of  light  the
             particles are not resolvable. Microscope counting
             is covered by British Standard 3406.
               Smaller particles. down to 0.001 pm diameter,
             can be examined using the electron microscope.
               The  two  major  disadvantages  of  microscopy
             are:  the  restricted  depth  of  focus  which  means
             that  examination  of  a  sample  with  a  wide  size
             range involves continual refocusing and the real
             possibility  of  missing  “out-of-focus”  particles
             during  a scan; it depends more than most  other
             methods on good representativeness.
               Several techniques are available for  the prep-
             aration  of  slides, the  all-important  factor being
             that  the sample is fully representative. The par-
             ticles  also need  to  be  well  separated  from  each
             other.  A  common  method  is  to  place  a  small
             fraction  of  the  sample onto  the  slide with  one   Figure 11.7  Examplesof eyepiecegraticules.
             drop  of  am  organic  fluid  such  as  methanol  or
             propanol  which  disperses  the  particles.  Subse-
             quent  evaporation  leaves  the  particles  suitably
             positioned.  The  fluid  obviously must  not  react   The slide is scanned in lateral strips, each strip
             with the particles, but it must have the ability to   an order or so wider than the largest particles. the
             “wet” them. Agitation with a soft brush can help.   objective being to  cover  the  whole  of  the  slide
             If  the particles have a sticky coating on them, it   area containing  particles. Typically one edge of
             may be necessary to remove this first by washing.   a chosen reference particle will be aligned with a
             One technique is to agitate, perhaps in water, to   major  graticule line using the longitudinal  stage
             allow ample time for all the particles to settle and   adjustment. The slide will then  be  traversed lat-
             then to pour  off  the fluid carefully, repeating as   erally along that line and all particles to the right
             necessary. Obviously  the  particles  must  not  be   of  that line will be counted, and measured using
             soluble in the fluid. Sometimes the material is first   the eyepiece scale. The slide will then be traversed
             agitated  in  a  fluid  and  then  one  drop  of  the   longitudinally to the right until the original par-
             particle-laden  fluid  transferred  to  the  slide.   ticle is in the same relative position but. for example,
             Representativeness  can  only  be  tested  by  the   five  major  graticule  lines  further  over  and  the
             repeatability of results from a number of samples.   counting process is repeated  for particles within
             Techniques  have  been  devised  for  transferring   the  strip formed  by  the  two  lines.  The  process
             samples from  suspension  onto films  within  the   involves  selecting  new  reference  particles  as
             suspension.  It  is  sometimes possible  to  collect   necessary. To avoid duplication, if  a particle lies
             samples  directly  onto  sticky  slides  coated  with   on one of the strip edge-lines, it is counted as if it
             grease, gelatin, or even rubber solution.   were  in the  strip to the right. Particles are allo-
               Earlier  methods  of  microscope  counting   cated to size bands suitably chosen to give. say, 10
             involved  the  use  of  an  optical  micrometer  by   points  on the  distribution curve. The  tedium  is
             which a cross-hair could be aligned with each side   relieved if operators work in pairs, one observing,
             of each particle in turn and the difference meas-   one recording. alternately.
             ured.  As  can  be  imagined  this  was  slow  and   Some graticules have been designed containing
             tedious  and  now  the  most  commonly  used   systems  of  opaque  and  open  circles,  the  sizes
             methods  involve calibrated  field graticules. The   arranged  in  various orders of  progression.  This
             graticules are engraved with a scale (Figure 11.7)   can  assist the  classification of  particles by  coni-
             nominally  divided,  for  example,  into  20 pm,   parison into size bands, each bounded by  one of
             100 pm, and 1 mm steps. Calibration is dependent   the circles.
             on  the  magnification  and  this  is  usually  finely   When sizing irregularly shaped particles, micro-
             adjustable  by  moving  the  objective  relative  to   scope  counting  introduces  a  bias  because  the
             the eyepiece slightly, or by adjusting the zoom if   particles tend  to lie  in their  most  stable  orient-
             available.  Calibration  is  effected  by  comparing   ation. By  making measurement of a distribution
             the field graticule with a stage graticule, similarly   of randomly orientated particles on a slide along
             and  accurately  engraved.  When  set,  the  stage   a fixed direction, one obtains a two-dimensional
             graticule is replaced by the sample slide.   statistical mean diameter.
   172   173   174   175   176   177   178   179   180   181   182