Page 380 - Instrumentation Reference Book 3E
P. 380

Potentiometry and specific ion measurement  363

                                   Center
                                   conductor
                                   (white)





                                                                                ; Toofened membranf
                           Shield
                           (black insulation                                    3  Cap araldited Io bung
                           over braided shield)                                 4  Aging Ci wire
                                                                                5  Tube
                                                                                6  Topbung
                                                                                7  2 pH filling
                                                                                8  7 pH filling
                                                                           -






                           Shielded
                           electrode cable
                           Glass-to-metal seal
                            ('"total glass" construction)
                                                                           -
                                                                              Min. amount of air
             Silver         Internal solution                                 to be trapped
             chloride       pH-sensitive membrane                          -4
             bead
             Figure 17.1  6  pH measuring electrode. Courtesy
            the Foxboro Company.
                                                      Figure 17.17  industrial electrode. Courtesy Kent
                                                      industrial Measurements Ltd. Analytical Instruments
             suspended  solids,  viscous  liquids,  or  reaction
            products  of  the process stream. Where this  is  a
            danger. the faster flow type of plug must be used.   mended  for low temperature  applications unless
             Many  rodtine  measurements  can,  however,  be   provision is made to prevent freezing or precipi-
            made with the non-flowing electrode, thus avoid-   tation  in the electrode and any associated hard-
             ing the necessity of refilling, or arranging a pres-   ware.
             surized  continuous  supply.  Flowing  types  of   When  materials  such as  sulfides: alkali  phos-
            junctions are usually required where an accuracy   phates or carbonates. which will react with silver,
             of h0.02 pH units (hl or 2 mV) is required, where   are  present  in  the  process  stream,  either  non-
             frequent  or  large  temperature  or  composition   flowing  electrodes,  or  electrodes  containing
             changes occur, or where the process fluid is such   potassium  chloride  at  lmol per  liter  should  be
             that it is prone to foul the junction.   used. The diffusion rate of silver can be neglected
              The temperature of operation will influence the   in  the  non-flowing  type,  and  the  solubility  of
             choice  of  concentration  of  the  filling solutions.   silver chloride in potassium chloride at a concen-
             Potassium chloride solution  having a concentra-   tration of  1 mol per liter is only 1 or 2 percent of
             tion of 4mol per liter saturates and starts to pre-   that in a solution at 4mol per liter.
             cipitate  solids  at  about  19"C,  and  will  freeze   High  temperatures with wide fluctuations are
             at -4  "G, while if the concentration  is reduced to   best  handled  by  potassium  chloride  solution  at
             1 mol  per  liter  the  solution  will  freeze at  -2 "C   1 mol per liter.
             without becoming saturated. Thus, no precipita-
             tion will take place in the solution of  lower con-
             centration. Although not damaging, precipitated   17.7.2  Measurement of pH
             potassium chloride and associated silver chloride
             will  tend  to  clog reference junctions  and  tubes,   Glass  electrode  Almost  all  pH  measurements
             decreasing electrolyte flow rate and increasing the   are best made with a glass electrode (the earliest
             risk  of  spurious  potentials.  For  these  reasons,   of  the  ion-selective electrodes), the  e.m.f.  being
             flowing  reference  electrodes  are  not  recom-   measured  relative  to  a  reference  electrode.  The