Page 384 - Instrumentation Reference Book 3E
P. 384

Potentiometry and specific ion measurement 367

             g  Gate
             d  Drain
             s  Source


















            Figure 17.23  pH measuring circuit using field effect transistors.

              If  the iso-potential point is arranged to be the   log meters have a facility of  expanding the scale
            locus  of  the  slope  of' the measuring  instrument,   so  that  the  precision  of  the  reading  can  be
            the  pH  measuring  circuit  can  be  modified  to   increased up to 10 times. Digital outputs are also
            include a temperature sensor arranged to change   available, with the most sensitive ones reading to
            the negative feedback so that the circuit compen-   0.001 pH unit (unlikely to be meaningful in prac-
            sates  for  the  change  in  slope  of  the  e.m.f./pH   tice) or 0.1 mV. Instruments incorporating micro-
            relationship.  It  is  important to  realize  that  the   processors  are  also  now  available-these   can
            temperature  compensation  only corrects  for the   calculate  the  concentration  of  substances  from
            change in the electrode response due to tempera-   pH measurements and give readout in concentra-
            ture change and the iso-potential  control setting   tion units.  Blank and volume corrections  can be
            therefore enables pH electrodes calibrated at one   applied automatically.
            temperature to be used at another. The iso-poten-
            tial  control  does  not  compensate for the  actual   Precision and accuracy  Measurements reprodu-
            change  in  pH  of  a  solution  with  temperature.   cible to 0.05 pH units are possible in well buffered
            Thus  if  pH  is  being  measured  to  establish  the   solutions  in  the pH  range  3  to  10.  For  routine
            composition  of  a  solution  one must  carry  out   measurements  it  is  rarely  possible  to  obtain  a
            the measurements at constant temperature.   reproducibility of better than kO.01 pH units.
              A few commercial pII meters have  a variable   In poorly buffered solutions reproducibility may
            iso-potential control so that they can be used with   be no better than &O.  1 pH unit and accuracy may
            several different combinations of electrodes but it   be lost by the absorption of carbon dioxide or by
            is  more  generally the  case that  pH meters  have   the presence of  suspensions, sols, and gels. How-
            fixed iso-potential control  settings and can  only   ever, measured  pH  values  can  often  be  used  as
            be  used  with  certain  combinations  of  pH  and   control  parameters  even  when  their  absolute
            reference electrodes. It is strongly recommended   accuracies are in doubt.
            that,  with  fixed  iso-potential  control  settings,
            both  the  glass  and  reference  electrodes  be   Sodium ion error  Glass electrodes for pH meas-
            obtained from the manufacturer of the pH meter.   urement  are  selective  for  hydrogen  ions,  not
            Temperature  compensation  circuits  generally   uniquely  responsive  to  them.  and  so  will  also
            work  only on the pH and direct  activity ranges   respond  to  sodium  and  other  ions  especially  at
            of a pH meter and not on the millivolt, expanded   alkaline  pH  values  (more than  about  11).  This
            millivolt, and relative millivolt ranges.   effect causes the pH value to be underestimated.
              Modern  pH  meters  with  analog  displays  are   Sodium ions produce the greatest error, lithium
            scaled 0 to 14pH units with the sniallest division   ions about a  half, potassium  ions about a fifth.
            on  the  scale  equivalent  to  0.1  unit  giving  the   and other ions less than a tenth of the error due to
            possibility  of  estimating  0.02pH units  by  inter-   sodium ions. One can either standardize the elec-
            polation.  The  millivolt  scale  is  generally  0  to   trode in an alkaline buffer solution  containing a
            1400mV  with  a  polarity  switch,  or  -700  to   suitable concentration of the appropriate salt, or
            +700mV  without  one.  The  smallest division  is   better,  use  the  special lithium  and  cesium  glass
            10mV, allowing estimation  to 2mV. Many ana-   electrodes developed for use in solutions of high
   379   380   381   382   383   384   385   386   387   388   389