Page 389 - Instrumentation Reference Book 3E
P. 389

372  Chemical analysis: electrochemical techniques

                                                     called “pIon meters.” Scales are provided, analog-
                                                     ous to the pH scale, for ions of various valencies
                                                     andlor  a  scale  that  can  be  calibrated  to  read
                                                     directly  in  terms  of  concentration  or  valency.
                                                     Meters  manufactured  as  pIon  meters  generally
                                                     also have pH and millivolt scales. To date, plon
                                                     scales only cover ions with charges of *1  and 52
                                                     because no ion-selective electrodes for determin-
                                                     ing ions of high charge are yet available commer-
                   13 cm long        13 cm long      cially. Direct activity scales read in relative units
                   3  mm diameter    S mm diameter
                                                     only and so must be calibrated before use in the
                                                     preferred measurement units.
                                                       As  with  pH  meters,  pIon  meters  can  be
                                                     obtained  with  analog  and  digital  displays, with
                                                     integral  microprocessors,  with  recorder  and
                                                     printer  outputs, and  with  automatic standardi-
                                                     zation. Temperature compensation can be incor-
                       ;kesilver                     porated  but although ion-selective and reference
                   ~
                        10 mm diameter   Platinum rod   electrode   combinations   have   iso-potential
                                     d-l   mm diameter   points,  the  facility  of  being  able to  set  the  iso-
                                                     potential  control  has  so  far  been  restricted  to
            Silver electrode   Platinum electrode    pH  measurement.  On dual pHlpIon meters  the
            Figure 17.29  Examples of metal redox electrodes.   iso-potential  control  (if  it  exists)  should  be
                                                     switched  out  on  the  pIon  and  activity  scales if
            reference electrodes  in  the  sample,  read  off  the   one wishes to make a slope correction when work-
            generated e.m.f. by means of a suitable measuring   ing with an ion-selective electrode at constant tem-
            circuit,  and obtain  the result  from a calibration   perature.
            curve  relating  e.m.f.  and  concentration  of  the   For  the  best  accuracy  and  precision  pIon
            substance being determined. The difference from   meters  should  be  chosen  that  can  discriminate
            pH determinations is that most ion-selective elec-   0.1 mV for direct potentiometry;  1 mV discrimin-
            trode  applications  require  the  addition  of  a   ation  is  sufficient when  using ion-selective elec-
            reagent  to buffer  or adjust  the ionic strength  of   trodes as indicators for titrimetric methods.
            the sample before measurement  of the potential.
            Thus,  unlike  measurement  of  pH  and  redox
            potentials,  ion-selective  electrodes  cannot  be   Practical  arrangements  For  accurate  poten-
            immersed  directly  in  a  plant  stream  of  liquid   tiometry, the temperature  of  the  solution being
            product,  and a  sampling arrangement  has to be   analyzed  and  the  electrode  assembly should  be
            used.  However  this  can  usually  be  done  quite   controlled and ideally all analyses should be car-
            simply.                                   ried out at the same temperature, e.g., by using a
                                                      thermostatically controlled water bath. Solutions
                                                      must  also be  stirred; otherwise the  e.m.f. devel-
            pHandpIon meters  High impedance e.m.f. meas-   oped by the electrode may not be representative
            uring circuits must be  used with most ion-selec-   of  the  bulk  of  the  solution.  A  wide  range  of
            tive electrodes and are basically the same as used   stirring speeds is  possible but  too  slow a  speed
            for measuring pH with a glass electrode. The pH   may  give  long  response  times  and  too  high  a
            meters measure  e.m.f. in  millivolts and are also   speed may generate heat in the solution. Precau-
            scaled in pH units. Provided the calibration con-   tions must also be taken  to minimize contamin-
            trol on the pH meter (which relates the measured   ation.
            e.m.f. to a fixed point on the pH scale) has a wide   Taking all these items into account, best results
            enough range of adjustment, the pH scale can be   in  the laboratory can  be  obtained  by mounting
            used for any univalent positive ion, for example,   the electrodes in a flow cell through which the test
            measurement  with  a  sodium-selective electrode   solution is being pumped;  see Figure  17.30. This
            can  be  read  on  the  meter  as  a  pNa  scale   is a mandatory arrangement for on-stream instru-
            (or  - log CN~). Measurements  with  electrodes   ments and in the laboratory in cases where the ion
            responding  to  divalent  or negative  ions  cannot   concentration  being  determined  is  close  to  the
            be  related  directly  to  the  pH  scale.  However,   limit of detection of the electrode.
            manufacturers generally make some modification   Flow cells should be constructed  of  a material
            to pH meters to simplify measurements with ion-   that will  not  contaminate a  sample with  the  ion
            selective electrodes, and the modified meters are   being  determined; the  flow  rates  of  the  solution
   384   385   386   387   388   389   390   391   392   393   394