Page 390 - Instrumentation Reference Book 3E
P. 390

Potentiometry and specific ion measurement 373
             must be high enough to provide “stirring” but low   Automatic  chemical  standardization  takes
             enough that sample volumes are kept low. There   place at preset intervals (in this case once every
             should be good displacement of a previous sample   24 hours) with provision for manual initiation of
             by  an incoming one, solution from  the reference   the  sequence  at  any  time.  The  standardization
             electrode should not reach the measuring electrode,   sequence commences by activating a valve to stop
             and  when  liquid  flow  through  a  flow  cell  stops,   the sample flow and to allow a sodium ion solu-
             the cell must retain liquid around the electrodes to   tion  of  known  strength  (the  standard  sodium
             prevent them drying out. Finally a flow cell should   solution) to be pumped into the flow cell. When
             be  waterjacketed  so  that  its  temperature can  be   the electrodes have stabilized in the new solution,
             controlled. Suitable flow cells can be machined out   the amplifier output is compared with a present
             of Perspex and are available commercially.   standard  value  in  the  auto-compensation  unit
               Pumps  used  must  be  capable  of  pumping  at   and  any  error  causes  a  servo-potentiometer  to
             least  two  channels  simultaneously  at  different   be driven so as to adjust the output signal to the
             rates,  the  larger volume  for the sample and the   required  value. The monitor  is then  returned  to
             lesser for the reagent  solution. Peristaltic pumps   measurement of the sample. The standardization
             are the most  frequently  used. It  follows that  all   period  lasts  30  minutes,  a warning  lamp  shows
             interconnecting  tubing and other components in   that  standardization  is  taking  place,  and  any
             contact with the sample must be Inert with respect   alarm and control contacts are disabled. It is also
             to the ion being determined.              possible  to  check  the  stability  of  the  amplifier
               As  direct potentiometric determination of ions   and,  by  a  manual  introduction  of  a  second
             by ion-selective electrodes requires more frequent   sodium  standard, to check and adjust  the  scale
             calibration  than  the  more  stable  pH  systems,   length.
             industrially developed ion-selective electrode sys-
             tems  often  incorporate  automatic  recalibration.
             This makes them more expensive than pH meas-   Conditioning and storage of  electrodes  The manu-
             uring systems. A typical scheme for an ion-selec-   facturer’s instructions regarding storage and pre-
             tive monitor (in this case for sodium) is shown in   treatment  of  electrodes  should  be  followed
             Figures  17.31 and 17.32.                 closely. The general rules are that (a) glass elec-
               Sampl’e water  flows to the constant  head unit   trodes should not be  allowed to dry out because
             and is then pumped  anaerobically  at a constant   reconditioning  may  not  be  successful, (b)  solid
             rate into the flow cell where it is equilibrated with   state electrodes can be stored in deionized water
             ammonia  gas obtained  by  pumping  a  stream of   for long periods, dry-covered with protective caps
             air  thro’ugh  ammonia  solution.  (Instead  of   and  generally  ready  for  use  after  rinsing  with
             ammonia gas a liquid amine could be  used, and   water,  (c)  gas-sensing  membranes  and  liquid
             this would then be the buffer liquid delivered by   ion-exchange  electrodes  must  never  be  allowed
             the  second  channel  of  the  pump.)  The  sample   to dry out, (d) reference electrodes are as import-
             then  flows through  the  flow cell to contact  the   ant  as  the  measuring  electrodes  and  must  be
             ion-selective and reference electrodes and then to   treated exactly as specified by the manufacturer.
             a drain.                                  The element must not  be allowed to dry out. as

                                                 Thermometer  Air   Electrode   E I ectrode
                                                 pocket   vent   pocket     pocket















                                                                                         I
              Figurel7.30  Flow cell for                          0            50
             ion-selective electrodes.                            I             I
   385   386   387   388   389   390   391   392   393   394   395