Page 395 - Instrumentation Reference Book 3E
P. 395

378  Chemical analysis: electrochemical techniques

          cathode and anode. Oxygen will then continue to
          be  reduced at the cathode, causing the flow of a
          current, the magnitude  of  which is proportional
          to the  partial  pressure  of  oxygen in  the  sample
          gas.
           The only materials in contact with the process
          are PVC and PTFE, and the membrane is recessed
          so that it does not suffer mechanical damage. The
          cell  needs  to be recharged with  a new  supply of
          electrolyte at 3- or 6-month intervals depending on      Yrconia tube
          the operating conditions, and the membrane can
          be replaced easily should it be damaged.
           The  cell  current  is  amplified by  a  solid  state
          amplifier,  which  gives  a  voltage  output  which
          can  be  displayed  on  a  selection  switch  giving   A I u m i n a
          ranges of 0-1,  0-5,  0-10  or 0-25  percent oxygen   tube
          and  a calibration  adjustment. The calibration is
          checked by using a reference gas, or air when the
          instrument  should  read  20.9  percent  oxygen  on
          the  0-25  percent  scale. The  instrument  has  an      Inner
          accuracy fl percent of scale range at the calibra-       electrode
          tion  temperature but  an error  of  f3 percent of
          the  reading  will  occur  for  a  16°C departure  in
          operating temperature.
            When in use the sensor may be housed in an in-
                                                                   Outer
          line  type housing  or in  a  dip-type  of  assembly,   ‘f   electrode
          usually made of PVC suitable for pressures up to   4 O2
                                                                     gas
                                                                Measured
          3.5 bar.                                     la)      Partial pressure P,
                                                   Figure 17.36  Oxygen probe. Courtesy Kent
                                                   Instruments.
          17.8.3  High temperature ceramic sensor oxygen
          probes
          Just as an electrical potential can be developed at   Thus by  positioning the probe in a stack or flue
          a glass membrane which is a function of the ratio   where the  temperature  is  above  600°C a  direct
          of the hydrogen concentrations  on either side, a   measurement of the oxygen present may be made.
          pure zirconia  tube maintained  at high  tempera-   (In another manufacturer’s instrument the probe
          ture will develop a potential between its surfaces   is  maintained  at a  temperature  of  850°C  by  a
          which  is  a  function  of  the  partial  pressure  of   temperature-controlled  heating  element.)  The
          oxygen which is in contact with its surfaces. This   instrument  illustrated  can  operate  from  600  to
          is  the  principle  involved  in  the  oxygen  meter   1200 “C, the reading being corrected for tempera-
          shown in Figure 17.36.                   ture, which is measured by  a thermocouple. The
            The potential developed is given by the Nernst   probe  is  protected  by  a  silicon  carbide  sheath.
          equation:                                The zirconia used is stabilized with calcium.
                                                     Standard  instruments  have  ranges  of  oxygen
          E, = (~~/4~){1n[internal partial pressure of  0;- ions]/   concentration  of  20.9-0.1  percent,  1000-1 ppm,
                     [external partial pressure of  0;- ions]}   10-5-10-25  partial  pressure  and  can  measure
                                                   oxygen with an accuracy of better than +IO  per-
          Thus, if the potential difference between the sur-   cent of the reading.
          faces is measured by  platinum electrodes in con-   As  temperatures  in  excess  of  600°C must  be
          tact with the two surfaces a measure may be made   used some of the oxygen in the sample will react
          of the ratio of the partial pressure of the oxygen   with  any  combustible  gas  present,  e.g.,  carbon
          inside and outside the probe. If dry instrument air   monoxide and hydrocarbons. Thus, the measure-
          (20.9 percent oxygen) is fed into the inside of the   ment will be lower than the correct value but will
          probe,  the partial  pressure of  oxygen inside the   still afford a rapid means of following changes in
          tube  may  be  regarded  as constant,  so  that  the   the  oxygen  content  of  a  flue  gas  caused  by
          electrical potential measured in a similar manner   changes in combustion conditions.
          to  that  adopted  in  pH  measurement  will  be  a   When using this form of oxygen cell, one must
          measure  of  the  concentration  of  the  oxygen  in   be  aware that one is measuring “net oxygen” as
          the  atmosphere  around  the  measuring  probe.   any  combustible  material  in  the  sample  stream
   390   391   392   393   394   395   396   397   398   399   400