Page 94 - Macromolecular Crystallography
P. 94

IN-HOUSE MACROMOLECULAR DATA COLLECTION  83


          Protocol 5.2 Data collection using the MAR345 and the MARCCD using the MARDTB

           1. Move the detector back to ≥200 mm to give clearance  separated at this distance. If not, drive the detector
          and to avoid damaging the face of the detector.  further out and record frames until they are.
           2. Move the back-stop out of the way. This is done using  9. From the oscillation and the newly derived distance,
          the ‘remote control’ unit next to the instrument.  calculate the orientation matrix using the strategy
           3. Check that ϕ (PHI) on the drum of the goniostat is set  programme within the MAR or by another software
             ◦
          to 0 . Check that χ (CHI) is set to zero. (Some cryocooling  package. Determine the start ϕ and the oscillation range
                             ◦
          users like to set χ = 60 or 70 ; the remote control unit  in order to collect the highest percentage of the total
          allows toggling CHI between 0 and 60 .)    obtainable data in the shortest time. Confirm that along ϕ
                                   ◦
           4. (a) Mount the ‘crystal and the goniometer’ onto the  there are no overlaps, otherwise  ϕ has to be decreased.
          goniostat. Centre the crystal and align it to the centre of the  10. Optimize the beam using the beam optimization
          ‘cross-wires’ using the small viewing screen (see Centring  features on the MARDTB.
          crystals above).                          11. Start collecting the data. Try and process the data ‘on
            (An option on the MARDTB is to have motorized x, y  the fly’ during collection and perform data reduction on it in
          translations built into the goniometer; z-translation along ϕ  order to see how well it merges. Also inspect the frames
          is always available. When x, y motors are present an IUCr,  closely during data collection, looking for crystal pathologies
          Huber goniometer is not used, but the crystal is mounted  such as spot splitting and changes in the resolution of
          directly onto the ϕ axis with a magnetic cap.)  crystal diffraction at different crystal orientations.
            (b) Replace the back-stop.
           5. Move the detector to 80 nm for the MAR345 or
          35 mm for the MARCCD165 (corresponding to a resolution  Useful points to note
          of 1.4 Ångstroms). Unless the crystal being studied is a DNA  For the MAR345 a distance of 100 mm (50 mm for
          crystal or an exceptionally well-diffracting protein then the  the MARCCD165) corresponds to a resolution of
          final observed is likely to be less than this.  1.54 Ångstrom when using Copper Kα radiation.
           6. Check the crystal diffraction by setting up to take a  Distance to resolution conversions can be calculated
            ◦
          0.5 oscillation (for around 2 min) by specifying the  using simple trigonometry by using the expressions
          required parameters in the file, namely crystal name,  (for zero detector swing):
          distance, and oscillation range. If the diffraction is weak,
          repeat the exposure with a longer time.    Resolution =  ∗  −1
                                                            ∗
           7. Index the image and look for tell tale pathologies in  1.54/(2 sin[0.5 tan  (172.5/x)]) for the MAR345
          the image, such as spot splitting and high spot mosaicity.  Resolution =  ∗  −1
                                                            ∗
          If the crystal looks ok, proceed.           1.54/(2 sin[0.5 tan  (82.5/x)]) for the MARCCD
           8. Determine the resolution limit of the crystal from this  where the wavelength is assumed to be that for Copper Kα,
          image and drive the detector to the distance such that the  namely 1.54(178) Ångstroms and x is the crystal to detector
          edge of the detector corresponds to this diffraction limit, i.e.  distance.
          the diffraction image should fill the detector once recorded  If no image appears initially then this may be because
          at this distance. Confirm that the diffraction spots are well  the erase lamp has blown; this may happen especially at
                                                     synchrotrons, where they see much higher use.



        of photographic film without the need for chem-  5.6.1 Raxis-IV ++
        ical processing and time-consuming scanning, as
                                                     Thedetectorconsistsof30 cmsquareplatesmounted
        was the case when film was used to record macro-
                                                     onto a flexible belt. Once exposed the plate is driven
        molecular X-ray data. Three image plate systems are
                                                     round to the back of the detector and sucked by
        now marketed by Rigaku MSC: the Raxis HR (with
                                                     vacuum onto the inside of a metal hemicylinder. The
        350-mm plates for high resolution work); the Raxis-
                                                     plate is read out by a revolving mirror which spins
        IV ++ ; and the Raxis-HTC (with three 300-mm image
                                                     the length of the axis of the cylinder and, while pro-
        plates and a very fast readout by the use of a dual
                                                     gressing incrementally, illuminates the plate with
        read head).
   89   90   91   92   93   94   95   96   97   98   99