Page 96 - Macromolecular Crystallography
P. 96

IN-HOUSE MACROMOLECULAR DATA COLLECTION  85


                                  Scintillator  Light barrier
                            Fiber optic taper
                            with a 2.8:1
                            reduction ratio
                        4096 × 4096 pixel
                        CCD array
                                                               Scattered
                                                                X-rays
                                                                        Protein
                                                                        crystal
                                                                              X-ray





                         Thermo-electric
                         (peltier) cooler
                    Controller
                    and ADC







                                                                          X-ray
                                                              Light










        Figure 5.4 Schematic diagram of a CCD detector. (Reproduced with permission from Nolting, Methods in Modern Biophysics, Springer
        Verlag, 2004.)



        by Hamlin, Nielsen, and Xuong) and consists of  abbreviation for Desk Top Beam line). This assem-
        four chips bonded together to make an active area  bly is supplied with variable apertures ranging
                2
        of 210 cm and currently used on many synchrotron  from 0 to 5 mm, with 2.5 mm resolution. It also
        beamlines around the world.                  incorporates built-in ion chambers with very wide
                                                     dynamic ranges, permitting very precise optimiza-
                                                     tion of the beam. Either detector (image plate or
        5.6.3 The MAR345 and MARCCD                  CCD) can be tilted in the vertical to a maximum 2θ
                                                         ◦
                                                     of 30 . For the MAR345, the detector can be driven
        Both the MAR345 and the MARCCD are sup-      from 75 to 425 mm and the CCD detector from 20 to
        plied with a MARDTB goniostat (DTB being an  370 mm using the MARDTB.
   91   92   93   94   95   96   97   98   99   100   101