Page 109 - Mechanical Engineers' Handbook (Volume 2)
P. 109

98   Bridge Transducers





























                                   Figure 21 Remote-location double-shunt calibration of bridge transducer.


                          In the zero calibration phase the Zero Cal switch is moved downward so excitation is re-
                          moved from the bridge. The sensitivity resistor (R  ) is concurrently placed across the
                                                                   SENS
                          bridge input terminals, simulating the power supply impedance to result in the same overall
                          system impedance encountered in the data circuit. Zero bridge transducer output is recorded.
                             The next calibration phase is the series phase. In the series calibration mode the two
                          switches in Fig. 22 are closed with the Zero Cal switch back in its original position, intro-
                          ducing R and R into the circuit. This removes power from one corner of the bridge and
                                 c1    c2
                          puts a calibration resistor R in series with the sensitivity resistor and one side of the bridge
                                               c1
                          output. The second calibration resistor becomes intermediate between the R  -to-R con-
                                                                                     SENS   c1
                          nection and excitation return. This second resistor, R , is selected to maintain the approxi-
                                                                   c2
                          mate equivalent bridge impedance across the excitation. The calibration circuit then
                          electrically simulates the bridge transducer. The value of R is determined experimentally,
                                                                         c1
                          corresponding to some measurand equivalent.














                                            Figure 22 Series calibration of bridge transducer.
   104   105   106   107   108   109   110   111   112   113   114