6 Resistance Bridge Transducer Measurement System Calibration  101




















                                     Figure 25 Oscillograph frequency response display.



                           model plus shunt resistor adapter) is a stable, complete model of the transducer, it can be
                           used to perform an absolute end-to-end system calibration and can be a valuable tool in
                           troubleshooting.
                              Several disadvantages are encountered when using the shunt resistor bridge as a cali-
                           bration tool. Since some transducers are hard to model, it is difficult to ensure that the bridge
                           is a representative model of the transducer under all conditions. Furthermore, a different
                           bridge model is required for each major transducer design.


































                                            Figure 26 Shunt resistor bridge.