Page 156 - Mechanical Engineers' Handbook (Volume 2)
P. 156

2 Thermocouples   145

                           difference across that length when the thermocouple is in service. The cold-worked region
                           is often very short, which makes it difficult to detect.
                              A similar drop in calibration was reported by Bentley and Morgan for PtRh versus Pt
                                                                                   10
                           thermocouples in response to cold work introduced by handling. They reported a drop of
                           about 0.4% in the cold-worked region. Annealing at 200 C did not entirely remove the drift
                           in the PtRh leg. It is difficult to identify this problem by subsequent recalibration, since the
                           region of partially degraded material may be placed in a uniform temperature zone during
                           recalibration and hence play no part in generating the signal under calibration conditions. If
                           a thermocouple is suspected of being inhomogeneous, it should be tested for homogeneity
                           along its entire length. If the test shows that the wire is homogeneous, then no calibration
                           is required, since the used portion of the wire is the same as the unused portion. If the wire
                           is not homogeneous, no recalibration can be of value because it will be impossible to place
                           the temperature gradient in the same location for calibration as it was for service. This
                           situation is described in Figs. 10–12.
                              Assume that the thermocouple was exposed to the unfavorable environment only near
                           the hot end, as shown in Fig. 10, and that both wires became less active as a result of the
                           reaction. The output will drop, as shown. If this defective thermocouple were placed in a
                           usual calibration facility, all of the affected material would be in the region of uniform
                           temperature. The emf would be generated entirely by the material near the entrance of the
                           furnace, which was never changed—the wires between points 1 and 2 and between 6 and 7
                           in Fig. 11. Under such conditions, a perfectly normal signal would be developed.
                              Figure 12 illustrates a test for homogeneity that can identify a defective thermocouple.
                           To test a thermocouple for inhomogeneity, clamp the junction at a constant temperature and
































                                                                 Figure 10 Temperature distribution along the
                                                                 thermocouple, in service, and the resulting output
                                                                 after deterioration of the wires. (Reproduced
                                                                 from Ref. 2, with permission.)
   151   152   153   154   155   156   157   158   159   160   161