Page 269 - Pipeline Rules of Thumb Handbook
P. 269

256     Pipeline Rules of Thumb Handbook


























                                                                Figure 5. Block diagram illustrates gas chromatograph oven.












              Figure 4. Bypass loop reduces sampling lag time.







                     Sample conditioning system                Figure 6. Separated components move through cell detectors
                                                               for quantifying sample.
           To obtain a representative sample of gas, a probe should be
         inserted into the middle third of the pipeline. This procedure
         is as essential for good sampling for on-line GCs as it is for
         spot or composite samples. A major difference, however, is the
         gas pressure is reduced to 15psig at the probe for transport to
                                               6
         the analyzer to ensure no heavy component (C +) is lost. With
         spot or composite samples, care is taken to maintain pipeline  controller then blocks the sample flow and allows it to
         conditions. Sometimes it is useful to use a sample bypass loop  equilibrate to atmospheric pressure. The sample is then swept
         to reduce sampling lag times, as shown in Fig. 4.     onto a set of carefully selected chromatographic columns by
           Note that filtering and liquid coalescing is accomplished, if  a stream of helium gas for sequential separation of each com-
         needed, to protect the integrity of the analyzer columns.  ponent in the natural gas mixture. A block diagram is shown
         Sometimes double block and bleed sample switching is used  in Fig. 5.
         to prevent cross-contamination by the standard or other  As each column does its specified separation job, it is
         sample streams.                                       switched out of the helium stream. The separated compo-
                                                               nents make their way through the thermal conductivity cell
                                                               detector where they can be quantified. An example of a detec-
                      Gas chromatograph oven                   tor output is shown in Fig. 6.
                                                                 These peaks must be gated and the area under each
           Once the gas enters the oven, a repeatable gas sample   peak integrated to be quantified, which is the job of the
         is measured by a sample inject valve. The chromatograph   chromatograph controller.
   264   265   266   267   268   269   270   271   272   273   274