Page 172 - Tandem Techniques
P. 172

Page 155

            The sampling apparatus [8] is a simple but ingenious device, that allows short bursts of sample to be
            placed on the column at any pre-selected time by simple valve operations. A diagram  showing the
            principle of the injection system is shown in Figure 4.20. The operational sequence of the valving
            system is extremely simple. The normal, non-sampling condition, is depicted on the right. The upper
            vacuum valve is open and the lower vacuum valve closed. Helium is fed into the center tube causing
            helium to flow through the column and forcing the sample flow to be diverted out through the top
            vacuum port. By closing the top vacuum valve and the helium supply valve, and simultaneously
            opening the lower vacuum valve, the sample contained in the helium flow from the thermal analyzer
            can be made to flow directly into the analyzer column (shown on the left). This flow is allowed to
            proceed for a prescribed time period that places a defined volume of sample on the column. The valves
            are then reversed to their original position, the sampling procedure arrested and the separation
            developed. The system ensures that the sample only passes through fused quartz tubing on its way to
            the column and, in doing so, does not come in contact with any valve material or pass through any
            valve. This eliminates the possibility of selective absorption or sample decomposition on the valve
            surfaces.
























                                                         Figure 4.21
                                                    The Tube Arrangement of
                                                  the Valveless Sample Injector
   167   168   169   170   171   172   173   174   175   176   177