Page 297 - Tandem Techniques
P. 297

Page 281

            apparatus is simpler, and probably less expensive to make than the fluorescence spectrometer, its
            performance is also severely limited in comparison.






























                                                         Figure 7.19
                                               The Resolution of Convoluted Peaks
                                    Reprinted with permission from K. Tenabe, M. Glick, B. Smith,
                                   E, Volgtman and J. D. Winefordner, Anal. Chem., 59(8)(1987)1124,
                                            Copyright 1987 American Chemical Society

            There are distinct disadvantages to the use of the traditional dispersive type monochromators for
            analyzing the fluorescent light produced from the sample. Either the separation must be arrested leaving
            the sample stationary in the sample cell, and the fluorescence spectrum obtained by the usual stop-start
            procedure, or the analyst must be content with measuring the fluorescent light at a single or narrow
            band of wavelengths. Although it is difficult to select the excitation light by any means other than the
            dispersion type monochromator, it certainly would be possible to utilize a diode array system to analyze
            the fluorescent light. Thus the use of a dispersion monochromator to select the wavelength of the
            excitation light, and a combination of the simple dispersion monochromator and the
   292   293   294   295   296   297   298   299   300   301   302