Page 300 - Tandem Techniques
P. 300

Page 284

            apparatus is shown in Figure 7.20. Light from a xenon lamp was focused onto the flow cell by two
            curved mirrors. Between the mirrors was an interference filter which selected the wavelength of the
            excitation light. To change the excitation light, the filter had to be removed and an appropriate
            replacement inserted. The fluorescent light was then focused by means of a plane and curved mirror
            onto the diode array. The diode array was cooled by a Peltier cooling unit. The liquid chromatographic
            system, that was used in conjunction with the apparatus, consisted of a Perkin Elmer Series 10 isocratic
            pump, a Rheodyne 7510 injector with a 20 µl sample loop, and an ODS column, 15 cm long, 4.6 mm
            I.D. packed with particles 10 µm in diameter.

            The test sample used was a mixture of polycyclic aromatic hydrocarbons, which was eluted isocratically
            with a solvent mixture of 80% acetonitrile and 20% water. The separation they obtained is shown in
            Figure 7.21, and it is seen that the apparatus functions quite well, but the sensitivity compared with the
            simple fluorescence detector is relatively low. It is also seen from the two-dimensional presentation in
            Figure 7.21 that good spectra can be obtained from partially resolved peaks and would certainly aid in
            confirming the identity of a specific solute.

            A Tandem Instrument that Monitors UV Absorption, Fluorescence and Luminescence

            In early 1989 Yappert et al. (7) described a tandem instrument involving a liquid chromatograph and a
            spectrometric assembly that monitors UV absorption, fluorescence and luminescence simultaneously. A
            diagram of the spectrometric system is shown in Figure 7.22, and it is seen that it is a fairly complicated
            arrangement. Pulsed light from a broad wavelength xenon lamp passes through an IR filter to the first
            half mirror. Light reflected at right angles, passes to a second half mirror, and the light that passes
            through the second half mirror is focused onto a photoelectric cell. The output from this photoelectric
            cell is used in a feedback system to control the intensity of the light emitted from the xenon lamp.

            The light that passes directly through the first half mirror is focused onto a monochromator that selects
            the wavelength of the light that is to act as
   295   296   297   298   299   300   301   302   303   304   305